Evaluation of Analytical Performance of Variant II Turbo HbA1c Analyzer According to Sigma Metrics.

作者信息

Bozkaya Giray, Uzuncan Nuriye, Bilgili Sibel, Demirezen Ozlem

机构信息

İzmir Bozyaka Egitim ve Arastirma Hastanesi, İzmir, Turkey.

出版信息

J Med Biochem. 2019 Mar 1;38(1):33-37. doi: 10.2478/jomb-2018-0014. eCollection 2019 Mar.

Abstract

BACKGROUND

Hemoglobin A1c, (HbA1c) which is the major constituent of glycated hemoglobin, has been used in the follow-up of retrospective glycemia for years and in the diagnosis of diabetes mellitus nowadays. Since the analytical performance of HbA1c should be high likewise all laboratory tests, various quality control measures are used. Sigma metrics is one of these measures and it is the combination of bias, precision and total allowable error that ensures a general evaluation of analytical quality. The aim of our study was to evaluate the analytical performance of Bio-Rad's Variant Turbo II HbA1c analyzer according to sigma metrics.

METHODS

Sigma levels were calculated using the data obtained from two levels of internal and 12 external quality control materials (Bio-Rad) of Variant II Turbo HbA1c analyzer according to σ = (TEa% - Bias%) / CV% formula.

RESULTS

The mean sigma levels for low and high quality control materials were found to be 3.0 and 4.1, respectively.

CONCLUSIONS

The annual mean analytical performance of Variant II Turbo HbA1c analyzer was found to be acceptable according to sigma metrics. In order to be sure of the difference in HbA1c results indicating the success or failure in treatment but not arise from analytical variation, it is thought that more stringent quality control measures should be applied to reach higher sigma levels.

摘要

背景

糖化血红蛋白的主要成分血红蛋白A1c(HbA1c)多年来一直用于回顾性血糖监测,如今也用于糖尿病的诊断。由于HbA1c的分析性能应与所有实验室检测一样高,因此采用了各种质量控制措施。西格玛指标就是其中之一,它是偏差、精密度和总允许误差的组合,可确保对分析质量进行全面评估。我们研究的目的是根据西格玛指标评估伯乐公司的Variant Turbo II HbA1c分析仪的分析性能。

方法

根据σ =(TEa%-偏差%)/CV%公式,使用从Variant II Turbo HbA1c分析仪的两个水平的内部和12个外部质量控制材料(伯乐公司)获得的数据计算西格玛水平。

结果

低质量控制材料和高质量控制材料的平均西格玛水平分别为3.0和4.1。

结论

根据西格玛指标,发现Variant II Turbo HbA1c分析仪的年度平均分析性能是可接受的。为了确定HbA1c结果的差异是表明治疗成功或失败而非由分析变异引起,认为应采用更严格的质量控制措施以达到更高的西格玛水平。