Karlsson L
Department of Toxicology, University Hospital, Linköping, Sweden.
J Chromatogr. 1987 Jul 3;417(2):309-17. doi: 10.1016/0378-4347(87)80124-x.
A method is presented for determination of delta 9-tetrahydrocannabinol-11-oic acid in urine by means of a fully automated liquid chromatographic system. Aliquots (200 microliter) of hydrolysed urine from prison inmates were directly injected onto a pre-column, followed by chromatography on two columns with different selectivity: CN and C8 columns. To obtain both greater selectivity and a low detection limit a twin-detector principle was used, consisting of both ultraviolet and electrochemical detection. Urine samples found to be positive with the EMIT cannabinoid were analysed, and the results were compared with those obtained from a well established gas chromatographic-mass spectrometric method. The precision of the method was 2.8% at a mean concentration of 85 ng/ml and 13.4% for 6 ng/ml of the acid. The detection limit was below 5 ng/ml.
本文介绍了一种通过全自动液相色谱系统测定尿液中Δ⁹ - 四氢大麻酚 - 11 - 酸的方法。将来自监狱囚犯的水解尿液等分试样(200微升)直接注入预柱,然后在两根具有不同选择性的色谱柱(氰基柱和C8柱)上进行色谱分析。为了获得更高的选择性和低检测限,采用了双检测器原理,包括紫外检测和电化学检测。对经酶免疫分析大麻素检测呈阳性的尿液样本进行分析,并将结果与成熟的气相色谱 - 质谱法所获结果进行比较。该方法在酸平均浓度为85纳克/毫升时的精密度为2.8%,在酸浓度为6纳克/毫升时为13.4%。检测限低于5纳克/毫升。