Hamlet Sean M, Haggerty Christopher M, Suever Jonathan D, Wehner Gregory J, Andres Kristin N, Powell David K, Charnigo Richard J, Fornwalt Brandon K
Department of Electrical and Computer Engineering, University of Kentucky, Lexington, KY, USA.
Department of Pediatrics, University of Kentucky, Lexington, KY, USA.
J Cardiovasc Magn Reson. 2017 Mar 1;19(1):25. doi: 10.1186/s12968-017-0338-6.
Left ventricular (LV) torsion is an important indicator of cardiac function that is limited by high inter-test variability (50% of the mean value). We hypothesized that this high inter-test variability is partly due to inconsistent breath-hold positions during serial image acquisitions, which could be significantly improved by using a respiratory navigator for cardiovascular magnetic resonance (CMR) based quantification of LV torsion.
We assessed respiratory-related variability in measured LV torsion with two distinct experimental protocols. First, 17 volunteers were recruited for CMR with cine displacement encoding with stimulated echoes (DENSE) in which a respiratory navigator was used to measure and then enforce variability in end-expiratory position between all LV basal and apical acquisitions. From these data, we quantified the inter-test variability of torsion in the absence and presence of enforced end-expiratory position variability, which established an upper bound for the expected torsion variability. For the second experiment (in 20 new, healthy volunteers), 10 pairs of cine DENSE basal and apical images were each acquired from consecutive breath-holds and consecutive navigator-gated scans (with a single acceptance position). Inter-test variability of torsion was compared between the breath-hold and navigator-gated scans to quantify the variability due to natural breath-hold variation. To demonstrate the importance of these variability reductions, we quantified the reduction in sample size required to detect a clinically meaningful change in LV torsion with the use of a respiratory navigator.
The mean torsion was 3.4 ± 0.2°/cm. From the first experiment, enforced variability in end-expiratory position translated to considerable variability in measured torsion (0.56 ± 0.34°/cm), whereas inter-test variability with consistent end-expiratory position was 57% lower (0.24 ± 0.16°/cm, p < 0.001). From the second experiment, natural respiratory variability from consecutive breath-holds translated to a variability in torsion of 0.24 ± 0.10°/cm, which was significantly higher than the variability from navigator-gated scans (0.18 ± 0.06°/cm, p = 0.02). By using a respiratory navigator with DENSE, theoretical sample sizes were reduced from 66 to 16 and 26 to 15 as calculated from the two experiments.
A substantial portion (22-57%) of the inter-test variability of LV torsion can be reduced by using a respiratory navigator to ensure a consistent breath-hold position between image acquisitions.
左心室(LV)扭转是心脏功能的一个重要指标,但受测试间高变异性(平均值的50%)限制。我们推测,这种高测试间变异性部分归因于连续图像采集期间屏气位置不一致,而使用呼吸导航器进行基于心血管磁共振(CMR)的左心室扭转量化可显著改善这一情况。
我们通过两种不同的实验方案评估测量左心室扭转时与呼吸相关的变异性。首先,招募17名志愿者进行带有刺激回波位移编码(DENSE)的CMR检查,其中使用呼吸导航器测量并强制所有左心室基底部和心尖部采集之间的呼气末位置变异性。从这些数据中,我们量化了在强制呼气末位置变异性存在和不存在时扭转的测试间变异性,这为预期的扭转变异性确定了一个上限。对于第二个实验(20名新的健康志愿者),分别从连续屏气和连续导航门控扫描(单一接受位置)中获取10对DENSE基底部和心尖部电影图像。比较屏气扫描和导航门控扫描之间扭转的测试间变异性,以量化自然屏气变化导致的变异性。为证明这些变异性降低的重要性,我们量化了使用呼吸导航器检测左心室扭转临床上有意义变化所需样本量的减少。
平均扭转角度为3.4±0.2°/cm。在第一个实验中,呼气末位置的强制变异性转化为测量扭转的相当大的变异性(0.56±0.34°/cm),而呼气末位置一致时的测试间变异性则低57%(0.24±0.16°/cm,p<0.001)。在第二个实验中,连续屏气的自然呼吸变异性转化为扭转变异性为0.24±0.10°/cm,这显著高于导航门控扫描的变异性(0.18±0.06°/cm,p = 0.02)。通过在DENSE检查中使用呼吸导航器,根据两个实验计算,理论样本量从66减少到16,从26减少到15。
通过使用呼吸导航器确保图像采集之间屏气位置一致,可减少左心室扭转测试间变异性的很大一部分(22 - 57%)。