Thaler Evangeline, Toledo Felippe, Korte Holger
Division for Health and Socials Studies in Physiotherapy, University of Applied Sciences Hochschule Fresenius, Munich, Germany.
Division for Health and Social Studies, University of Applied Sciences Hochschule Fresenius, Munich, Germany.
Geriatr Orthop Surg Rehabil. 2017 Mar;8(1):44-48. doi: 10.1177/2151458516681141. Epub 2017 Feb 15.
Although electrotherapy appears to have particularly interesting applications in the field of postoperative orthopedic rehabilitation, relatively little scientifically based research has been conducted in the area of electrotherapy with regard to safety involving patients with orthopedic implants.
Three electrotherapy forms were tested, such as high-volt stimulation (HVS), transcutaneous electric nerve stimulation (TENS), and galvanic current (GAL), using a model system containing a metal implant plate to evaluate whether heating in excess of 3°C would occur.
All changes in temperature for HVS, GAL, and TENS therapeutic electrical currents observed in our model system fall below the predefined 3°C.
To the best of our knowledge, this is the first experimental based observation that prolonged exposure to a direct electrical current at therapeutic strength does not result in heating of metal titanium plates.
尽管电疗法在术后骨科康复领域似乎有特别有趣的应用,但在电疗法涉及骨科植入物患者安全性方面,相对较少有基于科学的研究。
使用包含金属植入板的模型系统测试了三种电疗法形式,即高压刺激(HVS)、经皮电刺激神经疗法(TENS)和直流电疗法(GAL),以评估是否会出现超过3°C的发热情况。
在我们的模型系统中观察到的HVS、GAL和TENS治疗电流的所有温度变化均低于预先设定的3°C。
据我们所知,这是首个基于实验的观察结果,即长时间暴露于治疗强度的直流电不会导致金属钛板发热。