Suppr超能文献

基于荧光近红外发射的 aza-BODIPY 染料的痕量氨传感器。

Trace Ammonia Sensors Based on Fluorescent Near-Infrared-Emitting aza-BODIPY Dyes.

机构信息

Institute of Analytical Chemistry and Food Chemistry, Graz University of Technology , Stremayrgasse 9, 8010 Graz, Austria.

出版信息

Anal Chem. 2017 Mar 7;89(5):2859-2865. doi: 10.1021/acs.analchem.6b04045. Epub 2017 Feb 10.

Abstract

Highly sensitive ammonia sensors for environmental monitoring are presented. The sensing materials are based on fluorescent BF-chelated tetraarylazadipyrromethene dyes (aza-BODIPYs) dyes physically entrapped in polyurethane hydrogels and dispersed in silicone rubber. This layer is covered by a hydrophobic porous Teflon membrane used as an additional proton barrier and light scattering layer. The dual-lifetime referenced (DLR) sensors make use of near-infrared (NIR)-emitting Egyptian blue as a reference material and in combination with optical fibers are read-out via a compact phase-fluorometer. The detectable concentration range can be tuned by the choice of aza-BODIPY dye or/and the hydrogel matrix. The most sensitive sensor has a limit of detection (LOD) of 0.11 μg/L and the upper detectable concentration of 300 μg/L. No cross-sensitivity toward pH is observed. The sensors show remarkable operational stability with no noticeable drift over a period of 2 weeks.

摘要

用于环境监测的高灵敏度氨传感器。传感材料基于荧光 BF-螯合四芳基偶氮二吡咯甲川染料(aza-BODIPY 染料),物理包埋在聚氨酯水凝胶中并分散在硅橡胶中。该层由疏水性多孔聚四氟乙烯膜覆盖,用作附加的质子阻挡层和光散射层。双寿命参考(DLR)传感器利用近红外(NIR)发射的埃及蓝作为参考材料,并结合光纤通过紧凑型相荧光计进行读出。可通过选择 aza-BODIPY 染料或/和水凝胶基质来调整可检测浓度范围。最敏感的传感器的检测限(LOD)为 0.11 μg/L,可检测的上限浓度为 300 μg/L。对 pH 值没有交叉敏感性。传感器表现出显著的操作稳定性,在两周的时间内没有明显的漂移。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验