Berger Annemarie, Salla Sabine, Keppler Oliver T, Rabenau Holger F
Institute of Medical Virology, Hospital of the Goethe University of Frankfurt, Frankfurt/M., Germany.
Department of Ophthalmology, RWTH Aachen University, Germany.
Transfus Med Hemother. 2017 Jan;44(1):39-44. doi: 10.1159/000449207. Epub 2016 Nov 4.
The HCV RNA testing of potential cornea donors frequently relies on blood samples stored pre mortem. The recommended storage time of maximum 72 h frequently excludes a significant fraction of donors.
The influence of storage time of EDTA plasma samples at 4 °C on the viral load measured with the Roche HCV Quantitative Test vs. 2.0 was evaluated for 43 samples from HCV-positive individuals.
The mean reduction of the viral load after 4 °C storage for 6-8 days was 0.46 log IU/ml (range +0.17 to -1.66 log IU/ml). After 1-3 days a mean loss of 0.19 log IU/ml (range +0.30 to -1.41 log IU/ml) and after 3-5 days of 0.32 log IU/ml (range +0.36 to -1.81 log IU/ml) was observed. In 23.3% of samples, a viral load reduction ≥ 1 log IU/ml (1.0-1.81 log IU/ml) was found after prolonged storage (5-8 days). In none of the samples did the HCV load fall below the detection limit.
Plasma storage for up to 8 days can quantitatively reduce the HCV RNA load, yet has no influence on the reliability of a qualitative HCV RNA detection by this ultrasensitive test to determine the HCV status of serologically negative cornea donors.