Ziemann Christian, Stille Maik, Cremers Florian, Rades Dirk, Buzug Thorsten M
University Hospital Schleswig Holstein, Department of Radiotherapy / Campus Luebeck Department of Radiotherapy, Ratzeburger Allee 160, D-23562, Luebeck, Germany.
University of Luebeck, Institute of Medical Engineering, Ratzeburger Allee 160, D-23562, Luebeck, Germany.
J Appl Clin Med Phys. 2017 Jan;18(1):243-250. doi: 10.1002/acm2.12002. Epub 2016 Dec 5.
The quality of CT slices can be drastically reduced in the presence of high-density objects such as metal implants within the patients' body due to the occurrence of streaking artifacts. Consequently, a delineation of anatomical structures might not be possible, which strongly influences clinical examination.
The aim of the study is to clinically evaluate the retrieval of attenuation values and structures by the recently proposed Augmented Likelihood Image Reconstruction (ALIR) and linear interpolation in the presence of metal artifacts.
A commercially available phantom was equipped with two steel inserts. At a position between the metal rods, which shows severe streaking artifacts, different human tissue-equivalent inserts are alternately mounted. Using a single-source computer tomograph, raw data with and without metal rods are acquired for each insert. Images are reconstructed using the ALIR algorithm and a filtered back projection with and without linear interpolation. Mean and standard deviation are compared for a region of interest in the ALIR reconstructions, linear interpolation results, uncorrected images with metal rods, and the images without metal rods, which are used as a reference. Furthermore, the reconstructed shape of the inserts is analyzed by comparing different profiles of the image.
The measured mean and standard deviation values show that for all tissue classes, the metal artifacts could be reduced using the ALIR algorithm and the linear interpolation. Furthermore, the HU values for the different classes could be retrieved with errors below the standard deviation in the reference image. An evaluation of the shape of the inserts shows that the reconstructed object fits the shape of the insert accurately after metal artifact correction. Moreover, the evaluation shows a drop in the standard deviation for the ALIR reconstructed images compared to the reference images while reducing artifacts and keeping the shape of the inserts, which indicates a noise reduction ability of the ALIR algorithm.
HU values, which are distorted by metal artifacts, can be retrieved accurately with the ALIR algorithm and the linear interpolation approach. After metal artifact correction, structures, which are not perceptible in the original images due to streaking artifacts, are reconstructed correctly within the image using the ALIR algorithm. Furthermore, the ALIR produced images with a reduced noise level compared to reference images and artifact images. Linear interpolation results in a distortion of the investigated shapes and features remaining streaking artifacts.
由于条纹伪影的出现,患者体内存在金属植入物等高密度物体时,CT切片的质量会大幅下降。因此,可能无法清晰勾勒解剖结构,这对临床检查有很大影响。
本研究旨在临床评估在存在金属伪影的情况下,最近提出的增强似然图像重建(ALIR)和线性插值对衰减值和结构的恢复情况。
一个市售体模配备了两个钢质插入物。在金属棒之间出现严重条纹伪影的位置,交替安装不同的人体组织等效插入物。使用单源计算机断层扫描仪,针对每个插入物采集有和没有金属棒时的原始数据。使用ALIR算法以及带和不带线性插值的滤波反投影来重建图像。比较ALIR重建、线性插值结果、带有金属棒的未校正图像以及用作参考的没有金属棒的图像中感兴趣区域的均值和标准差。此外,通过比较图像的不同轮廓来分析插入物的重建形状。
测量的均值和标准差表明,对于所有组织类别,使用ALIR算法和线性插值可以减少金属伪影。此外,不同类别的HU值能够以低于参考图像标准差的误差恢复。对插入物形状的评估表明,经过金属伪影校正后,重建的物体与插入物的形状精确匹配。此外,评估显示与参考图像相比,ALIR重建图像的标准差有所下降,同时减少了伪影并保持了插入物的形状,这表明ALIR算法具有降噪能力。
通过ALIR算法和线性插值方法可以准确恢复因金属伪影而失真的HU值。经过金属伪影校正后,使用ALIR算法可在图像中正确重建因条纹伪影在原始图像中无法察觉的结构。此外,与参考图像和伪影图像相比,ALIR生成的图像噪声水平更低。线性插值会导致所研究形状的失真且仍存在条纹伪影。