An evaluation of three commercially available metal artifact reduction methods for CT imaging.

作者信息

Huang Jessie Y, Kerns James R, Nute Jessica L, Liu Xinming, Balter Peter A, Stingo Francesco C, Followill David S, Mirkovic Dragan, Howell Rebecca M, Kry Stephen F

机构信息

Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX 77030, USA. The University of Texas Health Science Center Houston, Graduate School of Biomedical Sciences, Houston, TX 77030, USA.

出版信息

Phys Med Biol. 2015 Feb 7;60(3):1047-67. doi: 10.1088/0031-9155/60/3/1047. Epub 2015 Jan 14.

Abstract

Three commercial metal artifact reduction methods were evaluated for use in computed tomography (CT) imaging in the presence of clinically realistic metal implants: Philips O-MAR, GE's monochromatic gemstone spectral imaging (GSI) using dual-energy CT, and GSI monochromatic imaging with metal artifact reduction software applied (MARs). Each method was evaluated according to CT number accuracy, metal size accuracy, and streak artifact severity reduction by using several phantoms, including three anthropomorphic phantoms containing metal implants (hip prosthesis, dental fillings and spinal fixation rods). All three methods showed varying degrees of success for the hip prosthesis and spinal fixation rod cases, while none were particularly beneficial for dental artifacts. Limitations of the methods were also observed. MARs underestimated the size of metal implants and introduced new artifacts in imaging planes beyond the metal implant when applied to dental artifacts, and both the O-MAR and MARs algorithms induced artifacts for spinal fixation rods in a thoracic phantom. Our findings suggest that all three artifact mitigation methods may benefit patients with metal implants, though they should be used with caution in certain scenarios.

摘要

对三种商业金属伪影减少方法进行了评估,以用于存在临床逼真金属植入物情况下的计算机断层扫描(CT)成像:飞利浦的O-MAR、通用电气使用双能CT的单色宝石光谱成像(GSI)以及应用了金属伪影减少软件(MARs)的GSI单色成像。通过使用多个体模评估每种方法,包括三个含有金属植入物(髋关节假体、补牙材料和脊柱固定棒)的仿真人体体模,评估指标包括CT数值准确性、金属尺寸准确性和条纹伪影严重程度降低情况。对于髋关节假体和脊柱固定棒的情况,所有三种方法都显示出不同程度的成功,而对于牙齿伪影,没有一种方法特别有效。还观察到了这些方法的局限性。当应用于牙齿伪影时,MARs低估了金属植入物的尺寸,并在金属植入物之外的成像平面引入了新的伪影,并且O-MAR和MARs算法在一个胸部体模中对脊柱固定棒都产生了伪影。我们的研究结果表明,所有三种伪影减轻方法可能对有金属植入物的患者有益,尽管在某些情况下应谨慎使用。

相似文献

引用本文的文献

5
The accuracy of Eclipse AXB and AAA dose algorithms with dental amalgam.带有牙科汞合金的Eclipse AXB和AAA剂量算法的准确性。
Phys Eng Sci Med. 2024 Dec;47(4):1571-1580. doi: 10.1007/s13246-024-01471-4. Epub 2024 Aug 14.

本文引用的文献

9
Dual-energy CT-based monochromatic imaging.基于双能 CT 的单能量成像。
AJR Am J Roentgenol. 2012 Nov;199(5 Suppl):S9-S15. doi: 10.2214/AJR.12.9121.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验