用于机械通气时支气管镜检查的柔性支气管镜原型的评估:概念验证研究。

Evaluation of a flexible bronchoscope prototype designed for bronchoscopy during mechanical ventilation: a proof-of-concept study.

机构信息

CHR Orléans, Service de Réanimation Polyvalente, Université F. Rabelais, Orléans, France.

Service de Réanimation Polyvalente, CHRU de Tours, Université F. Rabelais, Tours, France.

出版信息

Anaesthesia. 2017 Jun;72(6):719-728. doi: 10.1111/anae.13839. Epub 2017 Mar 14.

Abstract

Bronchoscopy during mechanical ventilation of patients' lungs significantly affects ventilation because of partial obstruction of the tracheal tube, and may thus be omitted in the most severely ill patients. It has not previously been possible to reduce the external diameter of the bronchoscope without reducing the diameter of the suction channel, thus reducing the suctioning capacity of the device. We believed that a better-designed bronchoscope could improve the safety of bronchoscopy in patients whose lungs were ventilated. We designed a flexible bronchoscope prototype with a drumstick-shaped head consisting of a long, thin proximal portion; a short and large distal portion for camera docking; and a large suction channel throughout the length of the device. The aims of our study were to test the impact of our prototype on mechanical ventilation when inserted into the tracheal tube, and to assess suctioning capacity. We first tested the efficiency of the suction channel, and demonstrated that the suction flow of the prototype was similar to that of conventional adult bronchoscopes. We next evaluated the consequences of bronchoscopy when using the prototype on minute ventilation and intrathoracic pressures during mechanical ventilation: firstly, in vitro using a breathing simulator; and secondly, in vivo using a porcine model of pulmonary ventilation. The insertion of adult bronchoscopes into the tracheal tube immediately impaired the protective ventilation strategy employed, whereas the prototype preserved it. For the first time, we have developed an innovative flexible bronchoscope designed for bronchoscopy during invasive mechanical ventilation, that both preserved the protective ventilation strategy, and enabled efficient suction flow.

摘要

在对患者肺部进行机械通气时,支气管镜检查会显著影响通气,因为这会导致气管导管部分阻塞,因此对于病情最严重的患者,可以省略该操作。以前,无法在不减小吸引通道直径的情况下减小支气管镜的外径,从而降低设备的吸引能力。我们认为,设计更好的支气管镜可以提高通气患者支气管镜检查的安全性。我们设计了一种带有鼓槌形头部的柔性支气管镜原型,由细长的近端部分、短而大的远端部分(用于摄像头对接)以及整个装置长度上的大吸引通道组成。我们研究的目的是测试将原型插入气管导管后对机械通气的影响,并评估吸引能力。我们首先测试了吸引通道的效率,并证明了原型的吸引流量与传统成人支气管镜相似。接下来,我们评估了在机械通气期间使用原型进行支气管镜检查对分钟通气量和胸腔内压力的影响:首先,在体外使用呼吸模拟器进行评估;其次,在体内使用猪的肺部通气模型进行评估。将成人支气管镜插入气管导管会立即损害所采用的保护性通气策略,而原型则可以保留该策略。我们首次开发了一种创新的柔性支气管镜,用于有创机械通气时的支气管镜检查,它既保留了保护性通气策略,又能实现有效的吸引流量。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验