Popot Marie-Agnes, Ho Emmie N M, Stojiljkovic Natali, Bagilet Florian, Remy Pierre, Maciejewski Pascal, Loup Benoit, Chan George H M, Hargrave Sabine, Arthur Rick M, Russo Charlie, White James, Hincks Pamela, Pearce Clive, Ganio George, Zahra Paul, Batty David, Jarrett Mark, Brooks Lydia, Prescott Lise-Anne, Bailly-Chouriberry Ludovic, Bonnaire Yves, Wan Terence S M

机构信息

Laboratoire des Courses Hippiques (LCH), 15 rue de Paradis, 91370, Verrières-le-Buisson, France.

Racing Laboratory, The Hong Kong Jockey Club, Sha Tin Racecourse, Sha Tin, NT, Hong Kong, China.

出版信息

Drug Test Anal. 2017 Sep;9(9):1400-1406. doi: 10.1002/dta.2191. Epub 2017 Jun 20.

Abstract

Cobalt is an essential mineral micronutrient and is regularly present in equine nutritional and feed supplements. Therefore, cobalt is naturally present at low concentrations in biological samples. The administration of cobalt chloride is considered to be blood doping and is thus prohibited. To control the misuse of cobalt, it was mandatory to establish an international threshold for cobalt in plasma and/or in urine. To achieve this goal, an international collaboration, consisting of an interlaboratory comparison between 5 laboratories for the urine study and 8 laboratories for the plasma study, has been undertaken. Quantification of cobalt in the biological samples was performed by inductively coupled plasma-mass spectrometry (ICP-MS). Ring tests were based on the analysis of 5 urine samples supplemented at concentrations ranging from 5 up to 500 ng/mL and 5 plasma samples spiked at concentrations ranging from 0.5 up to 25 ng/mL. The results obtained from the different laboratories were collected, compiled, and compared to assess the reproducibility and robustness of cobalt quantification measurements. The statistical approach for the ring test for total cobalt in urine was based on the determination of percentage deviations from the calculated means, while robust statistics based on the calculated median were applied to the ring test for total cobalt in plasma. The inter-laboratory comparisons in urine and in plasma were successful so that 97.6% of the urine samples and 97.5% of the plasma samples gave satisfactory results. Threshold values for cobalt in plasma and urine were established from data only obtained by laboratories involved in the ring test. Copyright © 2017 John Wiley & Sons, Ltd.

摘要

钴是一种必需的矿物质微量营养素,经常存在于马的营养和饲料补充剂中。因此,钴在生物样品中天然以低浓度存在。氯化钴的施用被认为是血液兴奋剂,因此被禁止。为了控制钴的滥用,必须建立血浆和/或尿液中钴的国际阈值。为实现这一目标,开展了一项国际合作,包括5个实验室参与的尿液研究和8个实验室参与的血浆研究的实验室间比对。生物样品中钴的定量分析采用电感耦合等离子体质谱法(ICP-MS)。环形试验基于对5个添加浓度范围为5至500 ng/mL的尿液样品和5个添加浓度范围为0.5至25 ng/mL的血浆样品的分析。收集、整理并比较不同实验室获得的结果,以评估钴定量测量的重现性和稳健性。尿液中总钴环形试验的统计方法基于计算平均值的百分比偏差的测定,而基于计算中位数的稳健统计方法应用于血浆中总钴的环形试验。尿液和血浆的实验室间比对取得成功,97.6%的尿液样品和97.5%的血浆样品给出了满意的结果。血浆和尿液中钴的阈值仅根据参与环形试验的实验室获得的数据确定。版权所有© 2017约翰·威利父子有限公司。