Max Planck Institute for Biological Cybernetics, Tuebingen, Germany.
IMPRS for Cognitive and Systems Neuroscience, Eberhard-Karls University of Tuebingen, Germany.
Magn Reson Med. 2018 Jan;79(1):529-540. doi: 10.1002/mrm.26658. Epub 2017 Mar 20.
To describe the process of calibrating a B shim system using high-degree (or high order) spherical harmonic models of the measured shim fields, to provide a method that considers amplitude dependency of these models, and to show the advantage of very high-degree B shimming for whole-brain and single-slice applications at 9.4 Tesla (T).
An insert shim with up to fourth and partial fifth/sixth degree (order) spherical harmonics was used with a Siemens 9.4T scanner. Each shim field was measured and modeled as input for the shimming algorithm. Optimal shim currents can therefore be calculated in a single iteration. A range of shim currents was used in the modeling to account for possible amplitude nonlinearities. The modeled shim fields were used to compare different degrees of whole-brain B shimming on healthy subjects.
The ideal shim fields did not correctly shim the subject brains. However, using the modeled shim fields improved the B homogeneity from 55.1 (second degree) to 44.68 Hz (partial fifth/sixth degree) on the whole brains of 9 healthy volunteers, with a total applied current of 0.77 and 6.8 A, respectively.
The necessity of calibrating the shim system was shown. Better B homogeneity drastically reduces signal dropout and distortions for echo-planar imaging, and significantly improves the linewidths of MR spectroscopy imaging. Magn Reson Med 79:529-540, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
描述使用测量的补偿场的高次(或高阶)球谐模型来校准 B 补偿系统的过程,提供一种考虑这些模型的幅度依赖性的方法,并展示在 9.4T(T)下全脑和单切片应用中非常高次 B 补偿的优势。
使用带有高达四阶和部分五/六阶(阶)球谐的插入补偿器和西门子 9.4T 扫描仪。每个补偿场都被测量并建模为补偿算法的输入。因此,可以在单个迭代中计算最佳补偿电流。在建模中使用了一系列补偿电流来考虑可能的幅度非线性。使用建模的补偿场来比较健康受试者的不同程度的全脑 B 补偿。
理想的补偿场不能正确补偿受试者的大脑。然而,使用建模的补偿场可以将 9 名健康志愿者的全脑 B 均匀度从 55.1(二阶)提高到 44.68 Hz(部分五/六阶),总应用电流分别为 0.77 和 6.8 A。
证明了校准补偿系统的必要性。更好的 B 均匀度可以大大减少回波平面成像的信号缺失和失真,并显著改善磁共振波谱成像的线宽。磁共振医学 79:529-540, 2018. © 2017 国际磁共振学会。