New-Generation Thromboelastography: Comprehensive Evaluation of Citrated and Heparinized Blood Sample Storage Effect on Clot-Forming Variables.

作者信息

Dias Joao D, Haney Elaine I, Mathew Blesy A, Lopez-Espina Carlos G, Orr Adrian W, Popovsky Mark A

机构信息

From the Department of Clinical Marketing, Haemonetics SA, Signy, Switzerland (Dr Dias); the Department of Scientific Research & Development, Haemonetics Corporation, Rosemont, Illinois (Mss Haney and Mathew and Mr Lopez-Espina); and the Department of Medical & Clinical Affairs, Haemonetics Corporation, Braintree, Massachusetts (Mr Orr and Dr Popovsky). Mr Orr is now with Anika Therapeutics Inc, Bedford, Massachusetts. Dr Dias and Ms Haney contributed equally to the design of this manuscript.

出版信息

Arch Pathol Lab Med. 2017 Apr;141(4):569-577. doi: 10.5858/arpa.2016-0088-OA. Epub 2017 Feb 28.

Abstract

CONTEXT

  • Thromboelastography (TEG) is a whole blood, real-time analyzer measuring the viscoelastic properties of the hemostasis process and allowing for individualized goal-directed therapy. However, routine use of TEG requires validation of sample storage effect on clot parameters.

OBJECTIVES

  • To establish the minimum time required for equilibration time and the maximum time for sample storage for all commercially available TEG tests for the new-generation TEG 6s and to determine how those times compare with the older generation TEG 5000.

DESIGN

  • Citrated and heparinized whole blood samples obtained from 20 healthy donors were analyzed for clot parameters at multiple time points for both the TEG 6s and the TEG 5000. Samples were activated with the citrated multichannel cartridge or the platelet-mapping cartridge in the TEG 6s or with recalcified kaolin in the TEG 5000.

RESULTS

  • All blood samples yielded TEG parameter results within reference ranges and had a tendency toward hypercoagulable profiles with increased storage time. Sample storage resulted in increased platelet inhibition with significant differences at 4 hours in the platelet-mapping cartridge (arachidonic acid percentage of inhibition, P = .002; adenosine diphosphate percentage of inhibition, P = .02).

CONCLUSIONS

  • For nonemergent cases or in a central laboratory setting, all tests provided reliable results for up to 4 hours in the citrated multichannel cartridge and for 3 hours for platelet function information in the platelet-mapping cartridge. In emergent/urgent situations in which the sample needs to be run immediately, RapidTEG and functional fibrinogen tests may be preferred.
摘要

背景

血栓弹力图(TEG)是一种全血实时分析仪,可测量止血过程的粘弹性特性,并允许进行个体化目标导向治疗。然而,TEG的常规使用需要验证样本储存对凝血参数的影响。

目的

确定新一代TEG 6s所有市售TEG检测的平衡时间所需的最短时间和样本储存的最长时间,并确定这些时间与旧一代TEG 5000相比如何。

设计

从20名健康供体获得的枸橼酸化和肝素化全血样本,在多个时间点对TEG 6s和TEG 5000的凝血参数进行分析。样本在TEG 6s中用枸橼酸化多通道试剂盒或血小板功能检测试剂盒激活,或在TEG 5000中用重新钙化的高岭土激活。

结果

所有血液样本的TEG参数结果均在参考范围内,且随着储存时间的延长有高凝倾向。样本储存导致血小板抑制增加,在血小板功能检测试剂盒中4小时时存在显著差异(花生四烯酸抑制百分比,P = 0.002;二磷酸腺苷抑制百分比,P = 0.02)。

结论

对于非紧急情况或在中心实验室环境中,在枸橼酸化多通道试剂盒中所有检测在长达4小时内提供可靠结果,在血小板功能检测试剂盒中血小板功能信息在3小时内可靠。在需要立即检测样本的紧急/急症情况下,可能首选快速TEG和功能性纤维蛋白原检测。