Suppr超能文献

术中生物人工瓣膜功能障碍导致严重二尖瓣反流。

Intraoperative Bioprosthetic Valve Dysfunction Causing Severe Mitral Regurgitation.

作者信息

Pang Philip Y K, Garwood Susan, Hashim Sabet W

机构信息

Section of Cardiac Surgery, Yale University School of Medicine, New Haven, Connecticut; Department of Cardiothoracic Surgery, National Heart Centre, Singapore.

Department of Anesthesiology, Yale University School of Medicine, New Haven, Connecticut.

出版信息

Ann Thorac Surg. 2017 Apr;103(4):e317-e319. doi: 10.1016/j.athoracsur.2016.09.019.

Abstract

Structural degeneration of bioprosthetic valves usually occurs gradually over time. Failure of a bioprosthetic valve immediately after implantation is extremely rare. Possible causes include obstruction of valve leaflets from preserved subvalvular tissue during chordal-sparing mitral valve replacement (MVR) or strut entrapment by suture loops. We report 2 cases of acute bioprosthetic mitral valve (MV) dysfunction involving newly implanted Perimount Theon (Edwards Lifesciences, Irvine, CA) bioprostheses, causing severe transvalvular mitral regurgitation (MR). Rapid diagnosis was achieved with the use of intraoperative transesophageal echocardiography (TEE). Operative assessment and examination of the explanted valves could not determine a definite cause of failure in either case.

摘要

生物人工瓣膜的结构退变通常会随着时间逐渐发生。生物人工瓣膜在植入后立即失效极为罕见。可能的原因包括在保留腱索的二尖瓣置换术(MVR)过程中,瓣膜小叶被保留的瓣下组织阻塞,或缝合环导致支柱卡压。我们报告2例急性生物人工二尖瓣(MV)功能障碍,涉及新植入的Perimount Theon(爱德华生命科学公司,加利福尼亚州欧文市)生物人工瓣膜,导致严重的二尖瓣反流(MR)。通过术中经食管超声心动图(TEE)实现了快速诊断。在这两例病例中,手术评估和对取出瓣膜的检查均无法确定明确的失效原因。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验