Biosens Bioelectron. 2017 Aug 15;94:584-588. doi: 10.1016/j.bios.2017.03.053. Epub 2017 Mar 27.
The determination of ethyl [4-oxo-8-(3-chlorophenyl)-4,6,7,8-tetrahydroimidazo[2,1-c][1,2,4]triazin-3-yl]acetate (ETTA), a new anticancer prodrug, using adsorptive stripping voltammetry (AdSV) was described for the first time. This method is based on adsorptive/reductive behaviour of ETTA at an in situ plated bismuth film electrode (BiFE) as a sensor. A number of experimental variables (e.g., a composition and pH of the supporting electrolyte, the conditions of bismuth film deposition, an accumulation potential and time, the scan rate, etc.) were thoroughly studied in order to achieve a high sensitivity. Experimental results under optimized conditions revealed an excellent linear correlation between the monitored voltammetric peak current and the ETTA concentration in the range of 2-50μgL following an accumulation time of 300s. The limit of detection (LOD) for ETTA following 300s of an accumulation time was 0.4μgL. The proposed facile, sensitive and inexpensive method was successfully applied to the determination of ETTA in serum. The investigated prodrug was extracted from serum using SPE method.
首次描述了采用吸附溶出伏安法(AdSV)测定新型抗癌前药乙基[4-氧代-8-(3-氯苯基)-4,6,7,8-四氢咪唑并[2,1-c][1,2,4]三嗪-3-基]乙酸酯(ETTA)。该方法基于 ETTA 在原位镀铋膜电极(BiFE)上的吸附/还原行为作为传感器。为了获得高灵敏度,对许多实验变量(例如,支持电解质的组成和 pH 值、铋膜沉积条件、累积电位和时间、扫描速率等)进行了深入研究。在优化条件下的实验结果表明,在 300s 的累积时间后,监测到的伏安峰电流与 ETTA 浓度之间存在极好的线性相关性,范围为 2-50μgL。在 300s 的累积时间后,ETTA 的检测限(LOD)为 0.4μgL。该简便、灵敏且廉价的方法成功应用于血清中 ETTA 的测定。该研究的前药采用 SPE 法从血清中提取。