van Es E M, Russell B C, Ivanov P, García Miranda M, Read D, Dirks C, Happel S
National Physical Laboratory, Hampton Road, Teddington, Middlesex TW11 UK.
Chemistry Department, University of Surrey, Guildford, Surrey GU2 7XH UK.
J Radioanal Nucl Chem. 2017;312(1):105-110. doi: 10.1007/s10967-017-5203-4. Epub 2017 Mar 13.
Accurate, low-level measurement of Ra in high volume water samples requires rapid pre-concentration and robust separation techniques prior to measurement in order to comply with discharge limits and drinking water regulations. This study characterises the behaviour of Ra and interfering elements on recently developed TK100 (Triskem International) extraction chromatography resin. Distribution coefficients over a range of acid concentrations are given, along with an optimised procedure that shows rapid pre-concentration and separation of Ra on TK100 resin is achievable for high volume (1 L) water samples without the need for sample pre-treatment.