Suppr超能文献

迭代重建和内核选择对肺密度定量 CT 测量的影响。

The effects of iterative reconstruction and kernel selection on quantitative computed tomography measures of lung density.

机构信息

Department of Medical Physics, University of Wisconsin School of Medicine and Public Health, Madison, WI, USA.

Department of Radiology, University of Wisconsin School of Medicine and Public Health, Madison, WI, USA.

出版信息

Med Phys. 2017 Jun;44(6):2267-2280. doi: 10.1002/mp.12255. Epub 2017 May 12.

Abstract

PURPOSE

To determine the effects of iterative reconstruction (IR) and high-frequency kernels on quantitative computed tomography (qCT) density measures at reduced X-ray dose.

MATERIALS AND METHODS

The COPDGene 2 Phantom (CTP 698, The Phantom Laboratory, Salem, NY) with four embedded lung mimicking foam densities (12lb, 20lb, and 4lb), as well as water, air, and acrylic reference inserts, was imaged using a GE 64 slice CT750 HD scanner in helical mode with four current-time products ranging from 12 to 100 mAs. The raw acquired data were reconstructed using standard (STD - low frequency) and Bone (high frequency) kernels with filtered back projection (FBP), 100% ASiR, and Veo reconstruction algorithms. The reference density inserts were manually segmented using Slicer3D (www.slicer.org), and the mean, standard deviation, and histograms of the segmented regions were generated using Fiji (http://fiji.sc/Fiji) for each reconstruction. Measurements of threshold values placed on the cumulative frequency distribution of voxels determined by these measured histograms at 5%, PD5 , and 15%, PD15 , (analogous to the relative area below -950 HU (RA-950) and percent density 15 (PD15) in human lung emphysema quantification, respectively), were also performed.

RESULTS

The use of high-resolution kernels in conjunction with ASiR and Veo did not significantly affect the mean Hounsfield units (HU) of each of the density standards (< 4 HU deviation) and current-time products within the phantom when compared with the STD+FBP reconstruction conventionally used in clinical applications. A truncation of the scanner reported HU values at -1024 that shifts the mean toward more positive values was found to cause a systematic error in lower attenuating regions. Use of IR drove convergence toward the mean of measured histograms (~100-137% increase in the number measured voxels at the mean of the histogram), while the combination of Bone+ASiR preserved the standard deviation of HU values about the mean compared to STD+FBP, with the added effect of improved spatial resolution and accuracy in airway measures. PD5 and PD15 were most similar between the Bone+ASiR and STD+FBP in all regions except those affected by the -1024 truncation artifact.

CONCLUSIONS

Extension of the scanner reportable HU values below the present limit of -1024 will mitigate discrepancies found in qCT lung densitometry in low-density regions. The density histogram became more sharply peaked, and standard deviation was reduced for IR, directly effecting density thresholds, PD5 and PD15 placed on the cumulative frequency distribution of each region in the phantom, which serve as analogs to RA-950 and PD15 typically used in lung density quantitation. The combination of high-frequency kernels (Bone) with ASiR mitigates this effect and preserves density measures derived from the image histogram. Moreover, previous studies have shown improved accuracy of qCT airway measures of wall thickness (WT) and wall area percentage (WA%) when using high-frequency kernels in combination with ASiR to better represent airway walls. The results therefore suggest an IR approach for accurate assessment of airway and parenchymal density measures in the lungs.

摘要

目的

确定迭代重建(IR)和高频核对降低 X 射线剂量时定量计算机断层扫描(qCT)密度测量的影响。

材料与方法

使用 COPDGene 2 体模(CTP 698,Phantom Laboratory,Salem,NY),该体模具有四个嵌入的肺模拟泡沫密度(12lb、20lb 和 4lb),以及水、空气和丙烯酸参考插件,在螺旋模式下使用 GE 64 切片 CT750 HD 扫描仪进行成像,电流-时间产品范围从 12 到 100 mAs。使用标准(STD-低频)和 Bone(高频)核、滤波反投影(FBP)、100%ASiR 和 Veo 重建算法对原始采集数据进行重建。使用 Slicer3D(www.slicer.org)手动对参考密度插件进行分割,使用 Fiji(http://fiji.sc/Fiji)为每个重建生成分割区域的平均值、标准差和直方图。还对阈值进行了测量,这些阈值是通过这些测量的直方图在 5%、PD5 和 15%、PD15 处(分别类似于人类肺气肿定量中的相对面积低于-950 HU(RA-950)和密度 15%(PD15))的累积频率分布确定的。

结果

与传统临床应用中常用的 STD+FBP 重建相比,在与 ASiR 和 Veo 一起使用高分辨率核时,每个密度标准品(<4 HU 偏差)和体模内电流-时间产品的平均 Hounsfield 单位(HU)均未受到显著影响。发现扫描仪报告的 HU 值截断在-1024 处,这导致平均 HU 值向正值偏移,从而导致在衰减较低的区域出现系统误差。IR 的使用使测量的直方图的平均值收敛(在直方图的平均值处测量的体素数量增加了 100-137%),而 Bone+ASiR 的组合与 STD+FBP 相比保留了 HU 值的标准差,同时还提高了气道测量的空间分辨率和准确性。在所有区域中,除了受-1024 截断伪影影响的区域外,Bone+ASiR 与 STD+FBP 之间的 PD5 和 PD15 最为相似。

结论

扩展扫描仪报告的 HU 值至目前的-1024 以下,将减轻低密度区域 qCT 肺密度计量中的差异。IR 使密度直方图的峰更加陡峭,标准偏差降低,这直接影响到每个体模区域累积频率分布上的密度阈值 PD5 和 PD15,这些阈值与通常用于肺密度定量的 RA-950 和 PD15 类似。高频核(Bone)与 ASiR 的结合减轻了这种影响,并保留了从图像直方图中得出的密度测量值。此外,先前的研究表明,在结合 ASiR 使用高频核时,气道壁厚度(WT)和壁面积百分比(WA%)的 qCT 气道测量的准确性得到了提高,这更好地代表了气道壁。因此,结果表明 IR 方法可用于准确评估肺部气道和实质密度测量值。

相似文献

引用本文的文献

本文引用的文献

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验