[Accuracy Verification of Respiratory-gated Radiotherapy that Combines the Respiration-Monitoring Device and Respiratory-gated System].

作者信息

Shintani Naoya, Monzen Hajime, Tamura Masaya, Asai Yoshiyuki, Shimomura Kouhei, Matsumoto Kenji, Okumura Masahiko, Nishimura Yasumasa

机构信息

Graduate School of Medicine, Kinki University Graduate School.

Department of Radiological Service, Kinki University Hospital.

出版信息

Igaku Butsuri. 2016;35(4):282-291. doi: 10.11323/jjmp.35.4_282.

Abstract

The purpose of this study is to evaluate the mechanical accuracy of a respiratory-gated radiation system that combines the Linear Indicator-equipped Abches respiration-monitoring device and the Varian Real-time Position Management system (LI-RPM system). This combined configuration, implemented for the first time in Japan, was compared with the stand-alone Varian RPM system (RPM system). The delay times, dose profiles, and output waveforms of the LI-RPM and RPM systems were evaluated using a self-produced dynamic phantom. The delay times for the LI-RPM and RPM systems were both 0.1 s for 4 s and 8 s test periods. The corresponding output waveform correlation factors (R) for the 4 s and 8 s test periods were 0.9981 and 0.9975, respectively. No difference was observed in the dose profiles of the two systems. Thus, the present results indicate that the proposed LI-RPM combined respiratory-gated radiation system has similar properties to the RPM system. However, it offers several advantages in terms of its versatility, including its alignment assistance capabilities for non-coplanar treatments.