Med Phys. 2018 Jan;45(1):74-80. doi: 10.1002/mp.12664. Epub 2017 Dec 3.
To evaluate the basic performance of a respiratory gating system using an Elekta linac and an Abches respiratory-monitoring device.
The gating system was comprised of an Elekta Synergy linac equipped with a Response gating interface module and an Abches respiratory-monitoring device. The latencies from a reference respiratory signal to the resulting Abches gating output signal and the resulting monitor-ion-chamber output signal were measured. Then, the flatness and symmetry of the gated beams were measured using a two-dimensional ionization chamber array for fixed and arc beams, respectively. Furthermore, the beam quality, TPR , and the output of the fixed gated beams were also measured using a Farmer chamber. Each of the beam characteristics was compared with each of those for nongated irradiation.
The full latencies at beam-on and beam-off for 6-MV gated beams were 336.4 ± 23.4 ms and 87.6 ± 7.1 ms, respectively. The differences in flatness between the gated and nongated beams were within 0.91% and 0.87% for the gun-target and left-right directions, respectively. In the same manner, the beam symmetries were within 0.68% and 0.82%, respectively. The percentage differences in beam quality and beam output were below 1% for a beam-on time range of 1.1-7 s.
The latency of the Elekta gating system combined with Abches was found to be acceptable using our measurement method. Furthermore, we demonstrated that the beam characteristics of the gating system using our respiratory indicator were comparable with the nongated beams for a single-arc gated beam delivery.
使用 Elekta 直线加速器和 Abches 呼吸监测设备评估呼吸门控系统的基本性能。
门控系统由配备响应门控接口模块的 Elekta Synergy 直线加速器和 Abches 呼吸监测设备组成。测量从参考呼吸信号到产生的 Abches 门控输出信号和产生的监测离子室输出信号的延迟。然后,使用二维电离室阵列分别测量固定和弧形束的门控束的平坦度和对称性。此外,还使用 Farmer 室测量束质量、TPR 和固定门控束的输出。将每个束特性与非门控照射的每个特性进行比较。
6-MV 门控束的全开和全关的总延迟分别为 336.4±23.4ms 和 87.6±7.1ms。门控和非门控束之间的平坦度差异分别在枪靶和左右方向为 0.91%和 0.87%。同样,束对称性分别为 0.68%和 0.82%。对于 1.1-7s 的束开启时间范围,束质量和束输出的百分比差异低于 1%。
使用我们的测量方法发现,Elekta 门控系统与 Abches 结合的延迟是可以接受的。此外,我们证明了使用我们的呼吸指示器的门控系统的束特性与单弧门控束输送的非门控束相当。