A preparative extraction step using disposable C18 low-pressure chromatography columns greatly improved the specificity of a commercially prepared digoxin radioimmunoassay (RIA). Elution solvents were isopropanol/water (15/85 by vol), which extracted most immunoreactive digitalis-like factors and metabolites, and methanol, which extracted digoxin for RIA. Many different digoxin RIA kits could be used. The coefficients of variation for replicates and duplicates were 4.6% and 5.2%. Analytical recoveries of digoxin standards in serum of 1.0, 0.5, and 0.1 microgram/L were 96%, 95%, and 88%, respectively. Serum digoxin was assayed by this method in 200 patients, 47 of whom were studied by HPLC-RIA. Values correlated better with "true digoxin" by HPLC-RIA (r = 0.93) than did values found by direct assay (r = 0.63). The mean for the isopropanol fraction as a percentage of the mean direct RIA value was higher for the 21 dialysis-dependent patients than was that found for the 179 nondialysis patients (P less than 0.004). The method is suggested as being most useful when metabolites or digitalis-like factors are known to be often high and values for digoxin are disproportionate to the dose.
使用一次性C18低压色谱柱的制备性提取步骤极大地提高了市售地高辛放射免疫分析(RIA)的特异性。洗脱溶剂为异丙醇/水(体积比15/85),其可提取大多数具有免疫反应性的洋地黄样因子和代谢产物,以及用于RIA提取地高辛的甲醇。可以使用许多不同的地高辛RIA试剂盒。重复检测和双份检测的变异系数分别为4.6%和5.2%。血清中1.0、0.5和0.1μg/L地高辛标准品的分析回收率分别为96%、95%和88%。用该方法对200例患者的血清地高辛进行了检测,其中47例患者通过高效液相色谱-放射免疫分析(HPLC-RIA)进行研究。与直接检测法(r = 0.63)相比,该方法检测的值与HPLC-RIA法测得的“真实地高辛”值相关性更好(r = 0.93)。21例依赖透析的患者异丙醇部分占直接RIA值平均值的百分比均值高于179例非透析患者(P < 0.004)。当已知代谢产物或洋地黄样因子含量通常较高且地高辛值与剂量不成比例时,建议使用该方法。