Gislason-Lee Amber J, Keeble Claire, Egleston Daniel, Bexon Josephine, Kengyelics Stephen M, Davies Andrew G
University of Leeds, Division of Biomedical Imaging, Leeds, United Kingdom.
University of Leeds, Division of Epidemiology and Biostatistics, Leeds, United Kingdom.
J Med Imaging (Bellingham). 2017 Apr;4(2):025501. doi: 10.1117/1.JMI.4.2.025501. Epub 2017 May 2.
This study aimed to determine whether a reduction in radiation dose was found for percutaneous coronary interventional (PCI) patients using a cardiac interventional x-ray system with state-of-the-art image enhancement and x-ray optimization, compared to the current generation x-ray system, and to determine the corresponding impact on clinical image quality. Patient procedure dose area product (DAP) and fluoroscopy duration of 131 PCI patient cases from each x-ray system were compared using a Wilcoxon test on median values. Significant reductions in patient dose ([Formula: see text]) were found for the new system with no significant change in fluoroscopy duration ([Formula: see text]); procedure DAP reduced by 64%, fluoroscopy DAP by 51%, and "cine" acquisition DAP by 76%. The image quality of 15 patient angiograms from each x-ray system (30 total) was scored by 75 clinical professionals on a continuous scale for the ability to determine the presence and severity of stenotic lesions; image quality scores were analyzed using a two-sample [Formula: see text]-test. Image quality was reduced by 9% ([Formula: see text]) for the new x-ray system. This demonstrates a substantial reduction in patient dose, from acquisition more than fluoroscopy imaging, with slightly reduced image quality, for the new x-ray system compared to the current generation system.
本研究旨在确定与当前一代X射线系统相比,使用具有最先进图像增强和X射线优化功能的心脏介入X射线系统的经皮冠状动脉介入治疗(PCI)患者的辐射剂量是否降低,并确定对临床图像质量的相应影响。使用Wilcoxon检验对中位数进行比较,对比了来自每个X射线系统的131例PCI患者病例的患者程序剂量面积乘积(DAP)和透视持续时间。新系统的患者剂量显著降低([公式:见原文]),透视持续时间无显著变化([公式:见原文]);程序DAP降低了64%,透视DAP降低了51%,而“电影”采集DAP降低了76%。75名临床专业人员对来自每个X射线系统的15例患者血管造影(共30例)的图像质量进行了连续评分,以确定狭窄病变的存在和严重程度;使用双样本[公式:见原文]检验分析图像质量评分。新X射线系统的图像质量降低了9%([公式:见原文])。这表明,与当前一代系统相比,新X射线系统的患者剂量大幅降低,采集剂量比透视成像剂量降低更多,图像质量略有下降。