Comprehensive assessment of patient image quality and radiation dose in latest generation cardiac x-ray equipment for percutaneous coronary interventions.

作者信息

Gislason-Lee Amber J, Keeble Claire, Egleston Daniel, Bexon Josephine, Kengyelics Stephen M, Davies Andrew G

机构信息

University of Leeds, Division of Biomedical Imaging, Leeds, United Kingdom.

University of Leeds, Division of Epidemiology and Biostatistics, Leeds, United Kingdom.

出版信息

J Med Imaging (Bellingham). 2017 Apr;4(2):025501. doi: 10.1117/1.JMI.4.2.025501. Epub 2017 May 2.

Abstract

This study aimed to determine whether a reduction in radiation dose was found for percutaneous coronary interventional (PCI) patients using a cardiac interventional x-ray system with state-of-the-art image enhancement and x-ray optimization, compared to the current generation x-ray system, and to determine the corresponding impact on clinical image quality. Patient procedure dose area product (DAP) and fluoroscopy duration of 131 PCI patient cases from each x-ray system were compared using a Wilcoxon test on median values. Significant reductions in patient dose ([Formula: see text]) were found for the new system with no significant change in fluoroscopy duration ([Formula: see text]); procedure DAP reduced by 64%, fluoroscopy DAP by 51%, and "cine" acquisition DAP by 76%. The image quality of 15 patient angiograms from each x-ray system (30 total) was scored by 75 clinical professionals on a continuous scale for the ability to determine the presence and severity of stenotic lesions; image quality scores were analyzed using a two-sample [Formula: see text]-test. Image quality was reduced by 9% ([Formula: see text]) for the new x-ray system. This demonstrates a substantial reduction in patient dose, from acquisition more than fluoroscopy imaging, with slightly reduced image quality, for the new x-ray system compared to the current generation system.

摘要

本研究旨在确定与当前一代X射线系统相比,使用具有最先进图像增强和X射线优化功能的心脏介入X射线系统的经皮冠状动脉介入治疗(PCI)患者的辐射剂量是否降低,并确定对临床图像质量的相应影响。使用Wilcoxon检验对中位数进行比较,对比了来自每个X射线系统的131例PCI患者病例的患者程序剂量面积乘积(DAP)和透视持续时间。新系统的患者剂量显著降低([公式:见原文]),透视持续时间无显著变化([公式:见原文]);程序DAP降低了64%,透视DAP降低了51%,而“电影”采集DAP降低了76%。75名临床专业人员对来自每个X射线系统的15例患者血管造影(共30例)的图像质量进行了连续评分,以确定狭窄病变的存在和严重程度;使用双样本[公式:见原文]检验分析图像质量评分。新X射线系统的图像质量降低了9%([公式:见原文])。这表明,与当前一代系统相比,新X射线系统的患者剂量大幅降低,采集剂量比透视成像剂量降低更多,图像质量略有下降。

相似文献

4
Patient radiation dose reduction using an X-ray imaging noise reduction technology for cardiac angiography and intervention.
Heart Vessels. 2016 May;31(5):655-63. doi: 10.1007/s00380-015-0667-z. Epub 2015 Apr 4.
5
Image noise reduction technology allows significant reduction of radiation dosage in cardiac device implantation procedures.
Pacing Clin Electrophysiol. 2017 Dec;40(12):1374-1379. doi: 10.1111/pace.13222. Epub 2017 Nov 19.
8
Latest-generation catheterization systems enable invasive submillisievert coronary angiography.
Herz. 2015 May;40 Suppl 3:233-9. doi: 10.1007/s00059-013-4015-8. Epub 2013 Dec 8.
10
Anthropometric parameters and radiation doses during percutaneous coronary procedures.
Phys Med. 2022 Aug;100:164-175. doi: 10.1016/j.ejmp.2022.06.013. Epub 2022 Jul 25.

引用本文的文献

1
Generative learning approach for radiation dose reduction in X-ray guided cardiac interventions.
Med Phys. 2022 Jun;49(6):4071-4081. doi: 10.1002/mp.15654. Epub 2022 Apr 18.
2
Radiation doses during cardiac catheterisation procedures in India: a multicentre study: Radiation dose study.
AsiaIntervention. 2020 Jul;6(1):25-33. doi: 10.4244/AIJ-D-18-00044. Epub 2020 Jul 20.
3
High filtration in interventional practices reduces patient radiation doses but not always scatter radiation doses.
Br J Radiol. 2021 Jan 1;94(1117):20200774. doi: 10.1259/bjr.20200774. Epub 2020 Nov 24.
4
An augmented reality system for image guidance of transcatheter procedures for structural heart disease.
PLoS One. 2019 Jul 1;14(7):e0219174. doi: 10.1371/journal.pone.0219174. eCollection 2019.

本文引用的文献

1
Methods for the analysis of ordinal response data in medical image quality assessment.
Br J Radiol. 2016 Jul;89(1063):20160094. doi: 10.1259/bjr.20160094. Epub 2016 Apr 12.
3
Patient radiation dose reduction using an X-ray imaging noise reduction technology for cardiac angiography and intervention.
Heart Vessels. 2016 May;31(5):655-63. doi: 10.1007/s00380-015-0667-z. Epub 2015 Apr 4.
6
Radiation dose in neuroangiography using image noise reduction technology: a population study based on 614 patients.
Neuroradiology. 2013 Nov;55(11):1365-72. doi: 10.1007/s00234-013-1276-0. Epub 2013 Sep 5.
8
Image noise reduction algorithm for digital subtraction angiography: clinical results.
Radiology. 2013 Nov;269(2):553-60. doi: 10.1148/radiol.13121262. Epub 2013 Jun 4.
9
Radiation cataract.
Ann ICRP. 2012 Oct-Dec;41(3-4):80-97. doi: 10.1016/j.icrp.2012.06.018. Epub 2012 Aug 22.
10
Patient radiation dose management in the follow-up of potential skin injuries in neuroradiology.
AJNR Am J Neuroradiol. 2013 Feb;34(2):277-82. doi: 10.3174/ajnr.A3211. Epub 2012 Aug 2.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验