Suppr超能文献

科学星期五下午:规划-07:一种用于三维辐射剂量测定的低扩散放射变色凝胶剂量计。

Sci-Fri PM: Planning-07: A low diffusion radiochromic gel dosimeter for three-dimensional radiation dosimetry.

作者信息

Babic S, Battista J, Jordan K

机构信息

Department of Medical Biophysics, The University of Western Ontario, London, ON.

London Regional Cancer Program - London Health Sciences Centre, London, ON.

出版信息

Med Phys. 2008 Jul;35(7Part3):3413. doi: 10.1118/1.2965979.

Abstract

PURPOSE

To develop a low diffusion radiochromic leuco crystal violet (LCV) hydrogel utilizing micelles, for three-dimensional (3-D) radiation dosimetry.

METHOD AND MATERIALS

Concentrations of LCV dye, Triton X-100 and trichloroacetic acid were varied to determine the optimal gel sensitivity for optical computed tomography (CT). Using a laser optical CT scanner at λ = 594 nm, diffusion rate measurements were performed on half-irradiated (6 MV x-rays) cuvette gel samples made with and without surfactant, respectively. A cylindrical 1 L gel volume was irradiated with a 12 MeV electron beam (Varian Clinac 2100C) to a dose of 30 Gy and scanned with cone- beam optical CT at λ ∼ 590 nm (Vista™, Modus Medical Devices Inc.).

RESULTS

The most radiation sensitive gel formulation was found to be: 1 mM LCV, 4 mM Triton X-100, 30 mM trichloroacetic acid and 4% gelatin. The diffusion rates of a LCV gel without and with surfactant present were about 2 and 20 times lower than the Fricke xylenol-orange gel system, respectively. Comparison of the central axis gel attenuation coefficients normalized at depth of maximum dose (d ) with TG21-corrected ion chamber data, were in agreement, thus, indicating energy and dose-rate independence.

CONCLUSION

Radiochromic LCV micelle gels show minimal diffusion effects and a dose response that is linear, energy and dose-rate independent. Optical CT scanned LCV micelle gels are a promising system for 3-D dose verification.

CONFLICT OF INTEREST

Two of the authors (JB, KJ) have a licensing agreement with Modus Medical Devices Inc. concerning the commercialization of Vista™.

摘要

目的

开发一种利用胶束的低扩散放射变色隐色结晶紫(LCV)水凝胶,用于三维(3-D)辐射剂量测定。

方法与材料

改变LCV染料、吐温X-100和三氯乙酸的浓度,以确定光学计算机断层扫描(CT)的最佳凝胶灵敏度。使用波长为λ = 594 nm的激光光学CT扫描仪,分别对添加和未添加表面活性剂制成的半辐照(6 MV X射线)比色皿凝胶样品进行扩散率测量。用12 MeV电子束(瓦里安Clinac 2100C)对1 L圆柱形凝胶体进行30 Gy的剂量照射,并在λ ∼ 590 nm下用锥束光学CT(Vista™,Modus Medical Devices Inc.)进行扫描。

结果

发现最具辐射敏感性的凝胶配方为:1 mM LCV、4 mM吐温X-100、30 mM三氯乙酸和4%明胶。存在和不存在表面活性剂时,LCV凝胶的扩散率分别比弗里克二甲苯酚橙凝胶系统低约2倍和20倍。将最大剂量深度(d )处归一化的中心轴凝胶衰减系数与经TG21校正的电离室数据进行比较,结果一致,因此表明与能量和剂量率无关。

结论

放射变色LCV胶束凝胶显示出最小的扩散效应以及线性、与能量和剂量率无关的剂量响应。光学CT扫描的LCV胶束凝胶是一种有前景的三维剂量验证系统。

利益冲突

两位作者(JB、KJ)与Modus Medical Devices Inc.就Vista™的商业化签订了许可协议。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验