Comparison between brain paste and polymer [125I]standards for quantitative receptor autoradiography.

作者信息

Davenport A P, Hall M D

机构信息

Parke-Davis Research Unit, Addenbrooke's Hospital Site, Cambridge, U.K.

出版信息

J Neurosci Methods. 1988 Aug;25(1):75-82. doi: 10.1016/0165-0270(88)90122-7.

Abstract

[125I]standards prepared from grey and white matter from rat brain have been compared to commercially available polymer-based standards using two types of radiation-sensitive film for use in quantitative receptor autoradiography. There was a linear relationship between the natural log of optical density and radioactivity (dpm/mm2) for both types of tissue standard exposed to Hyperfilm [3H] or betamax from 1-14 days. No difference between either the slopes or intercepts of these lines for each period of exposure was identified. There was also no significant difference between the slopes and intercepts of tissue-based standards and the lowest 5 activity levels of polymer standards. Over the complete range of ten radioactivity levels of polymer standards, the linear relationship was lost and a quadratic function was the best model for describing the relationship between 1n optical density and radioactivity. Moreover, a linear relationship between section thickness over the range frequently used in receptor studies (5-20 microns) was observed. These results suggest that the lower activity levels of polymer standards are comparable to tissue standards and can be used for quantitative receptor autoradiography.

摘要

已将由大鼠脑灰质和白质制备的[125I]标准品与市售的基于聚合物的标准品进行比较,使用两种用于定量受体放射自显影的放射敏感胶片。对于暴露于Hyperfilm [3H]或betamax 1至14天的两种组织标准品,光密度的自然对数与放射性(dpm/mm2)之间存在线性关系。在每个暴露时间段内,这些直线的斜率或截距均未发现差异。基于组织的标准品与聚合物标准品的最低5个活性水平之间的斜率和截距也没有显著差异。在聚合物标准品的十个放射性水平的整个范围内,线性关系消失,二次函数是描述1n光密度与放射性之间关系的最佳模型。此外,在受体研究常用的范围(5至20微米)内观察到切片厚度之间存在线性关系。这些结果表明,聚合物标准品的较低活性水平与组织标准品相当,可用于定量受体放射自显影。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验