SU-E-I-107: Suitability of Various Radiation Detectors Used in Radiation Therapy for X-Ray Dosimetry in Computed Tomography.

作者信息

Liebmann M, Poppe B, von Boetticher H

机构信息

Pius Hospital and Carl von Ossietzky University, Oldenburg.

Klinikum Links der Weser, Bremen.

出版信息

Med Phys. 2012 Jun;39(6Part5):3649-3650. doi: 10.1118/1.4734824.

Abstract

PURPOSE

Assessment of suitability for X-ray dosimetry in computed tomography of various ionization chambers, diodes and two-dimensional detector arrays primarily used in radiation therapy.

METHODS

An Oldelft X-ray simulation unit was used to irradiate PTW 60008, 60012 dosimetry diodes, PTW 23332, 31013, 31010, 31006 axial symmetrical ionization chambers, PTW 23343, 34001 plane parallel ionization chambers and PTW Starcheck and 2D-Array seven29 as well as a prototype Farmer chamber with a copper wall. Peak potential was varied from 50 kV up to 125 kV and beam qualities were quantified through half-value-layer measurements. Energy response was investigated free in air as well as in 2 cm depth in a solid water phantom and refers to a manufacturer calibrated PTW 60004 diode for kV-dosimetry.

RESULTS

The thimble ionization chambers PTW 31010, 31013, the uncapsuled diode PTW 60012 and the PTW 2D-Array seven29 exhibit an energy response deviation in the investigated energy region of approximately 10% or lower thus proving good usability in X-ray dosimetry if higher spatial resolution is needed or rotational irradiations occur.

CONCLUSIONS

It could be shown that in radiation therapy routinely used detectors are usable in a much lower energy region. The rotational symmetry is of advantage in computed tomography dosimetry and enables dose profile as well as point dose measurements in a suitable phantom for estimation of organ doses. Additional the PTW 2D-Array seven29 can give a quick overview of radiation fields in non-rotating tasks.

摘要

目的

评估主要用于放射治疗的各种电离室、二极管和二维探测器阵列在计算机断层扫描中进行X射线剂量测定的适用性。

方法

使用Oldelft X射线模拟装置对PTW 60008、60012剂量测定二极管、PTW 23332、31013、31010、31006轴对称电离室、PTW 23343、34001平行板电离室以及PTW Starcheck和2D-Array seven29,还有一个带铜壁的原型 Farmer 电离室进行辐照。峰值电位从50 kV变化到125 kV,并通过半值层测量对射束质量进行量化。在空气中以及在固体水模体2 cm深度处研究能量响应,并参考制造商校准的用于kV剂量测定的PTW 60004二极管。

结果

指形电离室PTW 31010、31013、未封装的二极管PTW 60012以及PTW 2D-Array seven29在研究的能量区域内表现出约10%或更低的能量响应偏差,因此如果需要更高的空间分辨率或进行旋转照射,在X射线剂量测定中证明具有良好的可用性。

结论

可以表明,在放射治疗中常规使用的探测器在低得多的能量区域也是可用的。旋转对称性在计算机断层扫描剂量测定中具有优势,并且能够在合适的模体中进行剂量分布以及点剂量测量以估计器官剂量。此外,PTW 2D-Array seven29可以在非旋转任务中快速给出辐射野的概况。