In recent years, public access defibrillation programs have exponentially increased the availability of automatic external defibrillators (AED) in public spaces but there are no data on their performance in our setting. We conducted a descriptive analysis of the performance of AED since the launch of a public defibrillation program in our region.
A retrospective analysis was conducted of electrocardiographic tracings and the performance of AED in a public defibrillation program from June 2011 to June 2015 in the province of Girona, Spain.
There were 231 AED activations. Full information was available on 188 activations, of which 82% corresponded to mobile devices and 18% to permanent devices. Asystole was the most prevalent rhythm (42%), while ventricular fibrillation accounted for 23%. The specificity of the device in identifying a shockable rhythm was 100%, but there were 8 false negatives (sensitivity 83%). There were 47 shockable rhythms, with a spontaneous circulation recovery rate of 49% (23 cases). There were no accidents related to the use of the device.
Nearly half of the recorded rhythms were asystole. The AED analyzed showed excellent safety and specificity, with moderate sensitivity. Half the patients with a shockable rhythm were successfully treated by the AED.
近年来,公众可获取除颤计划极大地增加了公共场所自动体外除颤器(AED)的可及性,但在我们的环境中尚无关于其性能的数据。我们对本地区启动公众除颤计划以来AED的性能进行了描述性分析。
对2011年6月至2015年6月西班牙赫罗纳省一项公众除颤计划中的心电图记录及AED性能进行回顾性分析。
AED共激活231次。188次激活有完整信息,其中82%对应移动设备,18%对应固定设备。心脏停搏是最常见的心律(42%),而室颤占23%。该设备识别可电击心律的特异性为100%,但有8例假阴性(敏感性83%)。有47次可电击心律,自主循环恢复率为49%(23例)。未发生与设备使用相关的事故。
记录的心律中近一半为心脏停搏。所分析的AED显示出极佳的安全性和特异性,敏感性中等。一半有可电击心律的患者经AED成功治疗。