腔内超声探头在宫颈癌近距离治疗中的应用:一种新型三维图像引导近距离治疗方法的模体研究

The use of trans-applicator intracavitary ultrasonography in brachytherapy for cervical cancer: phantom study of a novel approach to 3D image-guided brachytherapy.

作者信息

Tamaki Tomoaki, Miyaura Kazunori, Murakami Toshihiro, Kumazaki Yu, Suzuki Yoshiyuki, Nakano Takashi, Kato Shingo

机构信息

Department of Radiation Oncology, Fukushima Medical University, Fukushima, Fukushima.

Department of Radiation Oncology, Saitama Medical University International Medical Center, Hidaka, Saitama.

出版信息

J Contemp Brachytherapy. 2017 Apr;9(2):151-157. doi: 10.5114/jcb.2017.66588. Epub 2017 Mar 15.

Abstract

PURPOSE

To assess the feasibility of applying trans-applicator intracavitary ultrasonography to image-guided brachytherapy for cervical cancer.

MATERIAL AND METHODS

For this experiment, a phantom was created and included a polyethylene tube, intended to simulate a tandem applicator, which was inserted into chicken meat and embedded in agar, along with magnetic resonance imaging (MRI)-compatible ovoid applicators. Three-dimensional images of the phantom were obtained using computed tomography (CT), MRI (T2-weighted), and intracavitary ultrasonography sectional images acquired at 1 mm slice intervals. Intracavitary ultrasonography images were acquired from within the simulated tandem applicator using a radial transducer. Magnetic resonance imaging and intracavitary ultrasonography images were manually registered onto CT images. The chicken meat was contoured as the target volume independently on the CT, MRI, and intracavitary ultrasonography images, and the Dice similarity coefficient was used to compare the target volumes. The dose distributions of a sample brachytherapy plan were also evaluated.

RESULTS

Computed tomography, MRI, and intracavitary ultrasonography all visualized the three-dimensional phantom volumes. Intracavitary ultrasonography images depicted the meat with high echoic signals and a border clearly distinguishable from the surrounding agar. The Dice similarity coefficient values for the target volumes on CT vs. MRI, CT vs. intracavitary ultrasonography, and MRI vs. intracavitary ultrasonography were 0.966, 0.965, and 0.971, respectively, indicating similar contouring with the three modalities. Among the modalities, the differences in D, D, D, and D values were 1.8%, 2.9%, 3.7%, and 2.9%, respectively.

CONCLUSIONS

Three-dimensional reconstructed trans-applicator intracavitary ultrasonographic images clearly depicted meat tissue within the phantom, and could thus be used for brachytherapy planning. This study proves the concept of trans-applicator intracavitary ultrasonography for intracavitary brachytherapy. Further research such as development of intracavitary ultrasonography system, 3D reconstruction method, ultrasonography-compatible applicators, and ultrasonography-based target concept is warranted to assess the potential clinical application.

摘要

目的

评估将经施源器腔内超声检查应用于宫颈癌图像引导近距离放射治疗的可行性。

材料与方法

在本实验中,制作了一个模型,其中包括一根聚乙烯管,用于模拟串联施源器,将其插入鸡肉中并嵌入琼脂中,同时还有与磁共振成像(MRI)兼容的卵圆形施源器。使用计算机断层扫描(CT)、MRI(T2加权)以及以1毫米层厚间隔获取的腔内超声断层图像来获得模型的三维图像。使用径向换能器从模拟串联施源器内部获取腔内超声图像。将磁共振成像和腔内超声图像手动配准到CT图像上。在CT、MRI和腔内超声图像上分别独立勾画出鸡肉作为靶区体积,并使用骰子相似系数来比较靶区体积。还评估了一个样本近距离放射治疗计划的剂量分布。

结果

CT、MRI和腔内超声均能显示模型的三维体积。腔内超声图像显示鸡肉具有高回声信号,且边界与周围琼脂明显可区分。CT与MRI、CT与腔内超声、MRI与腔内超声的靶区体积骰子相似系数值分别为0.966、0.965和0.971,表明三种模态下的轮廓勾画相似。在这些模态中,D、D、D和D值的差异分别为1.8%、2.9%、3.7%和2.9%。

结论

经施源器腔内超声的三维重建图像清晰显示了模型内的肌肉组织,因此可用于近距离放射治疗计划。本研究证明了经施源器腔内超声用于腔内近距离放射治疗的概念。有必要开展进一步的研究,如腔内超声系统的开发、三维重建方法、与超声兼容的施源器以及基于超声的靶区概念等,以评估其潜在的临床应用价值。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/fedd/5437077/1a9b7402a57f/JCB-9-29638-g001.jpg