Crum E M, O'Connor W J, Van Loo L, Valckx M, Stannard S R
a School of Sport and Exercise , Massey University , Palmerston North , New Zealand.
b Haagsche Hogeschool , Den Haag , Zuid-Holland , Netherlands.
Eur J Sport Sci. 2017 Sep;17(8):1037-1043. doi: 10.1080/17461391.2017.1330899. Epub 2017 May 30.
The Moxy is a novel, cutaneously placed muscle oxygen monitor which claims to measure local oxygen saturation (SmO) and total haemoglobin (THb) using near-infrared spectroscopy. If shown to be reliable, its data storage and telemetric capability will be useful for assessing localised O usage during field-based exercise. This study investigated the reliability of the Moxy during cycling and assessed the correlations between its measurements, whole-body O consumption (VO) and heart rate (HR).
Ten highly trained cyclists performed an incremental, step-wise cycling protocol on two occasions while wearing the Moxy. SmO, THb, VO and HR were recorded in the final minute of each five-minute stage. Data were analysed using Spearman's Order-Rank Coefficient (SROC), Intraclass Correlation (ICC), and Coefficient of Variance (COV). Significance was set at p ≤ .05.
SmO showed a 'strong' or 'very large' correlation between trials (SROC: r = 0.842-0.993, ICC: r = 0.773-0.992, p ≤ .01) and was moderately correlated with VO and HR (r = -0.71-0.73, p ≤ .01). SmO showed a moderate to high reliability at low intensities, but this decreased as relative exercise intensity increased. THb showed poor correlations between tests and with the other measured variables, but was highly reliable at all power outputs.
The Moxy is a reliable device to measure SmO at low to moderate intensities, but at higher intensities, greater variation in measurements occurs, likely due to tissue ischaemia or increased movement artefacts due to more frequent muscular contractions. THb has low variation during exercise, and does not appear to be a valid indicator of muscle oxygenation.
Moxy是一种新型的经皮放置的肌肉氧监测仪,据称可使用近红外光谱法测量局部氧饱和度(SmO)和总血红蛋白(THb)。如果被证明是可靠的,其数据存储和遥测功能将有助于评估野外运动期间的局部氧利用情况。本研究调查了Moxy在骑行过程中的可靠性,并评估了其测量值、全身氧消耗(VO)和心率(HR)之间的相关性。
10名训练有素的自行车运动员在佩戴Moxy的情况下,分两次进行递增式、逐步的骑行方案。在每五分钟阶段的最后一分钟记录SmO、THb、VO和HR。使用斯皮尔曼等级相关系数(SROC)、组内相关系数(ICC)和变异系数(COV)对数据进行分析。显著性设定为p≤0.05。
SmO在试验之间显示出“强”或“非常大”的相关性(SROC:r = 0.842 - 0.993,ICC:r = 0.773 - 0.992,p≤0.01),并且与VO和HR呈中度相关(r = -0.71 - 0.73,p≤0.01)。SmO在低强度时显示出中度至高可靠性,但随着相对运动强度的增加而降低。THb在测试之间以及与其他测量变量之间显示出较差的相关性,但在所有功率输出下都具有高度可靠性。
Moxy是一种在低至中等强度下测量SmO的可靠设备,但在较高强度下,测量值的变化更大,这可能是由于组织缺血或由于更频繁的肌肉收缩导致的运动伪影增加。THb在运动期间变化较小,似乎不是肌肉氧合的有效指标。