Portable near-infrared spectroscopy devices allow measurements of muscle oxygen saturation (SmO) in real time and non-invasively. To use NIRS for typical applications including intensity control and load monitoring, the day-to-day variability needs to be known to interpret changes confidently. This study investigates the absolute and relative test-retest reliability of the Moxy Monitor and investigates side differences of SmO at the vastus lateralis muscle of both legs in cyclists.
Twelve trained cyclists and triathletes completed 3 incremental step tests with 5 min step duration starting at 1.0 W/kg with an increase of 0.5 W/kg separated by 2-7 days. SmO was averaged over the last minute of each stage. For all power outputs, the intra-class coefficient (ICC), the standard error of measurement (SEM) and the minimal detectable change (MDC) were calculated. Dominant and non-dominant leg SmO were compared using a three-factor ANOVA and limits of agreement (LoA).
ANOVA showed no significant systematic differences between trials and side. For both legs and all intensities, the ICC ranged from 0.79 to 0.92, the SEM from 5 to 9% SmO and the MDC from 14 to 18% SmO. The bias and LoA between both legs were -2.0% ± 19.9% SmO.
Relative reliability of SmO was numerically good to excellent according to current standards. However, it depends on the specific analytical goal whether the test-retest reliability is deemed sufficient. Wide LoA indicate side differences in muscle oxygenation during exercise unexplained by leg dominance.
便携式近红外光谱仪设备可实时、无创地测量肌肉氧饱和度(SmO2)。为了将 NIRS 用于包括强度控制和负荷监测在内的典型应用,需要了解日常变化,以便有信心地解释变化。本研究调查了 Moxy 监测仪的绝对和相对测试-重测可靠性,并研究了自行车运动员双腿股外侧肌 SmO2 的侧差异。
12 名经过训练的自行车运动员和三项全能运动员完成了 3 次递增台阶测试,每次持续 5 分钟,起始功率为 1.0 W/kg,每 2-7 天增加 0.5 W/kg。在每个阶段的最后 1 分钟内平均 SmO2。对于所有功率输出,计算了组内系数(ICC)、测量误差(SEM)和最小可检测变化(MDC)。使用三因素 ANOVA 和一致性界限(LoA)比较优势和非优势腿 SmO2。
ANOVA 显示试验和侧位之间没有显著的系统差异。对于双腿和所有强度,ICC 范围为 0.79 至 0.92,SEM 为 5 至 9% SmO2,MDC 为 14 至 18% SmO2。双腿之间的偏差和 LoA 为-2.0%±19.9% SmO2。
根据当前标准,SmO2 的相对可靠性在数值上是良好到优秀的。然而,测试-重测可靠性是否被认为足够,取决于特定的分析目标。宽 LoA 表明在运动过程中肌肉氧合存在无法用腿优势解释的侧差异。