Abstract

OBJECTIVES

This study prospectively evaluates objective image quality (IQ), subjective IQ, and PI-RADS scoring of prostate MRI at 3.0T (3T) and 1.5T (1.5T) within the same patients.

METHODS

Sixty-three consecutive patients (64±9years) were prospectively included in this non-inferiority trial, powered at 80% to demonstrate a ≤10% difference in signal-to-noise (SNR) and contrast-to-noise ratio (CNR) of T2-weighted and diffusion-weighted imaging (T2WI, DWI) at 1.5T compared to 3T. Secondary endpoints were analysis of subjective IQ and PI-RADS v2 scoring.

RESULTS

All patients received multi-parametric prostate MRI on a 3T (T2WI, DWI, DCE) and bi-parametric MRI (T2WI, DWI) on a 1.5T scanner using body coils, respectively. SNR and CNR of T2WI were similar at 1.5T and 3T (p=0.7-1), but of DWI significantly lower at 1.5T (p<0.01). Subjective IQ was significantly better at 3T for both, T2WI and DWI (p<0.01). PI-RADS scores were comparable for both field strengths (p=0.05-1). Inter-reader agreement was excellent for subjective IQ assessment and PI-RADS scoring (k=0.9-1).

CONCLUSION

Prostate MRI at 1.5T can reveal comparable objective image quality in T2WI, but is inferior to 3T in DWI and subjective IQ. However, similar PI-RADS scoring and thus diagnostic performance seems feasible independent of the field strength even without an endorectal coil.

摘要

目的

本研究前瞻性评估同一患者在3.0T(3T)和1.5T(1.5T)场强下前列腺MRI的客观图像质量(IQ)、主观IQ及PI-RADS评分。

方法

本非劣效性试验前瞻性纳入63例连续患者(64±9岁),检验效能为80%,以证明1.5T与3T相比,T2加权成像和扩散加权成像(T2WI、DWI)的信噪比(SNR)和对比噪声比(CNR)差异≤10%。次要终点为分析主观IQ和PI-RADS v2评分。

结果

所有患者分别在3T扫描仪上接受多参数前列腺MRI(T2WI、DWI、DCE)检查,在1.5T扫描仪上使用体线圈接受双参数MRI(T2WI、DWI)检查。T2WI的SNR和CNR在1.5T和3T时相似(p=0.7-1),但DWI的SNR和CNR在1.5T时显著更低(p<0.01)。3T时T2WI和DWI的主观IQ均显著更好(p<0.01)。两种场强下的PI-RADS评分具有可比性(p=0.05-1)。阅片者间主观IQ评估和PI-RADS评分的一致性良好(k=0.9-1)。

结论

1.5T场强下的前列腺MRI在T2WI中可显示相当的客观图像质量,但在DWI和主观IQ方面不如3T。然而,即使不使用直肠内线圈,类似的PI-RADS评分及诊断性能似乎与场强无关。