Binny Diana, Mezzenga Emilio, Lancaster Craig M, Trapp Jamie V, Kairn Tanya, Crowe Scott B
Cancer Care Services, Royal Brisbane and Women's Hospital, Brisbane, Australia; Queensland University of Technology, Brisbane, Australia.
Medical Physics Unit, Istituto Scientifico Romagnolo per lo Studio e la Cura dei Tumori (IRST) IRCCS, Meldola, Italy.
Phys Med. 2017 Jun;38:105-110. doi: 10.1016/j.ejmp.2017.05.052. Epub 2017 May 24.
The aims of this study were to investigate machine beam parameters using the TomoTherapy quality assurance (TQA) tool, establish a correlation to patient delivery quality assurance results and to evaluate the relationship between energy variations detected using different TQA modules. TQA daily measurement results from two treatment machines for periods of up to 4years were acquired. Analyses of beam quality, helical and static output variations were made. Variations from planned dose were also analysed using Statistical Process Control (SPC) technique and their relationship to output trends were studied. Energy variations appeared to be one of the contributing factors to delivery output dose seen in the analysis. Ion chamber measurements were reliable indicators of energy and output variations and were linear with patient dose verifications.
本研究的目的是使用螺旋断层放疗质量保证(TQA)工具研究机器射束参数,建立与患者放疗质量保证结果的相关性,并评估使用不同TQA模块检测到的能量变化之间的关系。获取了两台治疗机器长达4年期间的TQA每日测量结果。对射束质量、螺旋和静态输出变化进行了分析。还使用统计过程控制(SPC)技术分析了与计划剂量的偏差,并研究了它们与输出趋势的关系。在分析中,能量变化似乎是导致放疗输出剂量变化的因素之一。电离室测量是能量和输出变化的可靠指标,并且与患者剂量验证呈线性关系。