Berger Terry A
SFC Solutions Inc., 9435 Downing St., Englewood, FL 34224, USA.
J Chromatogr A. 2017 Aug 11;1510:82-88. doi: 10.1016/j.chroma.2017.06.021. Epub 2017 Jun 10.
The performance of a 3×50mm, 1.6μm d column with an immobilized polysaccharide stationary phase (ChiralPak IA-U) was evaluated for efficiency, and pressure drop, with respect to flow rate and modifier concentration using supercritical fluid chromatography (SFC). This appears to be the first such report using such a column in SFC. A unique low dispersion (ultra-high performance) SFC was used for the evaluation. The minimum reduced plate height of 2.78, indicates that the maximum efficiency was similar to or better than coated polysaccharide columns. Selectivity was different from ChiralPak AD, with the same chiral selector, as reported by many others. At high flows and high methanol concentrations, pump pressures sometimes approached 600bar. With 5% methanol, pressure vs. flow rate was non-linear suggesting turbulent flow in the connector tubing. The optimum flow rate (F) at 40% methanol was ≈0.8mL/min, where the column efficiency was highest. At 5% methanol, F increased to ≈1.6mL/min, but efficiency degraded noticeably. The differences in F suggests that the solute diffusion coefficients are a strong function of modifier concentration. Several sub-1min separations, including a 7.5s separation, are presented.
使用超临界流体色谱法(SFC),针对流速和改性剂浓度,对具有固定化多糖固定相(ChiralPak IA-U)的3×50mm、1.6μm内径柱的柱效和压降进行了评估。这似乎是首篇在SFC中使用此类色谱柱的相关报告。评估中采用了独特的低扩散(超高效)SFC。最小折合板高为2.78,这表明其最大柱效与涂覆多糖色谱柱相当或更优。如其他许多人所报道的,在使用相同手性选择剂的情况下,其选择性与ChiralPak AD不同。在高流速和高甲醇浓度下,泵压有时接近600bar。在甲醇含量为5%时,压力与流速呈非线性关系,这表明连接管中存在湍流。在甲醇含量为40%时,最佳流速(F)约为0.8mL/min,此时柱效最高。在甲醇含量为5%时,F增加至约1.6mL/min,但柱效显著下降。F的差异表明溶质扩散系数是改性剂浓度的强函数。文中给出了几次分离时间小于1分钟的分离结果,包括一次7.5秒的分离。