Bauer Westley S, Richardson Kelly A, Adams Nicholas M, Ricks Keersten M, Gasperino David J, Ghionea Simon J, Rosen Mathew, Nichols Kevin P, Weigl Bernhard H, Haselton Frederick R, Wright David W
Department of Chemistry, Vanderbilt University, Nashville, Tennessee 37235, USA.
Department of Biomedical Engineering, Vanderbilt University, Nashville, Tennessee 37235, USA.
Biomicrofluidics. 2017 Jun 2;11(3):034115. doi: 10.1063/1.4984788. eCollection 2017 May.
Rapid diagnostic tests (RDTs) designed to function at the point of care are becoming more prevalent in malaria diagnostics because of their low cost and simplicity. While many of these tests function effectively with high parasite density samples, their poor sensitivity can often lead to misdiagnosis when parasitemia falls below 100 parasites/l. In this study, a flow-through pipette-based column was explored as a cost-effective means to capture and elute more histidine-rich protein II (HRPII) antigen, concentrating the biomarker available in large-volume lysed whole blood samples into volumes compatible with -specific RDTs. A systematic investigation of immobilized metal affinity chromatography divalent metal species and solid phase supports established the optimal design parameters necessary to create a flow-through column incorporated into a standard pipette tip. The bidirectional flow inherent to this format maximizes mixing efficiency so that in less than 5 min of sample processing, the test band signal intensity was increased up to a factor of twelve from HRPII concentrations as low as 25 pM. In addition, the limit of detection per sample was decreased by a factor of five when compared to the RDT manufacturer's suggested protocol. Both the development process and commercial viability of this application are explored, serving as a potential model for future applications.
由于成本低且操作简单,旨在即时检测的快速诊断测试(RDTs)在疟疾诊断中越来越普遍。虽然这些测试中的许多对高寄生虫密度样本能有效发挥作用,但当寄生虫血症低于100个寄生虫/微升时,其低灵敏度往往会导致误诊。在本研究中,探索了一种基于流通式移液器的柱体,作为一种经济高效的方法来捕获和洗脱更多富含组氨酸的蛋白II(HRPII)抗原,将大量裂解全血样本中可用的生物标志物浓缩到与特定RDTs兼容的体积中。对固定化金属亲和色谱二价金属种类和固相支持物的系统研究确定了创建一个整合到标准移液器吸头中的流通柱所需的最佳设计参数。这种形式固有的双向流动使混合效率最大化,因此在不到5分钟的样品处理时间内,测试带信号强度从低至25皮摩尔的HRPII浓度增加到了12倍。此外,与RDT制造商建议的方案相比,每个样品的检测限降低了五倍。本研究探讨了该应用的开发过程和商业可行性,为未来应用提供了一个潜在模型。