Huckle P D, Summers M P
J Pharm Pharmacol. 1985 Oct;37(10):722-5. doi: 10.1111/j.2042-7158.1985.tb04950.x.
The influence of size, configuration and positioning of die wall strain gauges on the measurement of radially transmitted stress developed during tableting was investigated. Calibration of strain gauges attached to a cutaway die wall was achieved by compression of rubber-like materials in the die, Breon Polyblend 504 being more effective than red rubber for this purpose. Hysteresis of response observed when calibrating a thin-walled die was possible due to excessive distortion of the weakened die wall although the extent of this hysteresis varied with calibration material. The use of full bridge arrangements produced an increase in response when the compression site was moved away from the gauges. The opposite effect was seen when half bridge arrangements, using two active gauges, were used. The paradoxical effect observed when full bridge arrangements were used was shown to be due to straining of the compensating gauges. The dependence of die wall response on compact position was substantially reduced by the use of multiple gauges mounted along the die length and connected together to form a single gauge on each side of the die.
研究了冲模壁应变片的尺寸、形状和位置对压片过程中径向传递应力测量的影响。通过在模具中压缩类橡胶材料来实现对安装在剖切冲模壁上的应变片的校准,在此方面,Breon Polyblend 504比红色橡胶更有效。在校准薄壁模具时观察到的响应滞后现象,可能是由于弱化的模具壁过度变形所致,尽管这种滞后的程度随校准材料而变化。当压缩部位远离应变片时,采用全桥配置会使响应增加。当使用采用两个有源应变片的半桥配置时,观察到相反的效果。使用全桥配置时观察到的矛盾效应表明是由于补偿应变片的应变所致。通过沿模具长度安装多个应变片并将它们连接在一起,在模具的每一侧形成一个单一的应变片,模具壁响应对应压实位置的依赖性大大降低。