da Silva Marcelo Felix, de Oliveira Harley Francisco, Borges Leandro Frederich, Carrara Helio Humberto Angotti, Farina Jayme Adriano
Ann Plast Surg. 2018 Jan;80(1):67-70. doi: 10.1097/SAP.0000000000001177.
The purpose of this research was to develop an experimental model of dosimetry using a breast phantom and evaluate the effects of the metallic port in tissue expanders on dose distribution in postmastectomy radiotherapy. Dose distribution was assessed using an experimental acrylic dosimetry simulator in the absence and presence of a metallic disc (MD), which is similar to that used in tissue expanders containing a magnetic port, by collecting coronal and sagittal radiochromic films. Dosimetry film analysis did not show any changes in dose distribution, except for a MD shadow in the sagittal film where the dose distribution rate was on average 14% lower than in other areas. This model very closely resembled actual breast reconstruction with a tissue expander containing a magnetic port. Scattering or attenuation of the irradiation dose due to interference of the MD was not evident in areas that could jeopardize the effectiveness of radiation therapy. Therefore, the presence of the MD inside the tissue expander does not likely have an impact on radiotherapy effectiveness during immediate breast reconstruction.
本研究的目的是使用乳房模型开发一种剂量测定实验模型,并评估组织扩张器中的金属端口对乳房切除术后放疗剂量分布的影响。通过收集冠状面和矢状面放射变色胶片,使用实验性丙烯酸剂量测定模拟器在不存在和存在金属盘(MD)的情况下评估剂量分布,该金属盘类似于用于含有磁性端口的组织扩张器中的金属盘。剂量测定胶片分析未显示剂量分布有任何变化,除了矢状面胶片上的MD阴影,其剂量分布率平均比其他区域低14%。该模型与含有磁性端口的组织扩张器进行的实际乳房重建非常相似。在可能危及放射治疗效果的区域,未发现由于MD的干扰而导致的照射剂量散射或衰减。因此,组织扩张器内MD的存在不太可能对即刻乳房重建期间的放射治疗效果产生影响。