Pavoni Juliana Fernandes, Lizar Jessica Caroline, Borges Leandro Frederiche, Nicolucci Patricia, Krutman Yanai, Baffa Oswaldo
Department of Physics, Faculty of Philosophy, Sciences and Letters of Ribeirão Preto, University of São Paulo, Ribeirão Preto 14040-901, SP, Brazil.
Radiotherapy Sector, Clinical Hospital of the Ribeirão Preto Medical School, University of São Paulo, Ribeirão Preto 14048-900, SP, Brazil.
Gels. 2025 Apr 30;11(5):335. doi: 10.3390/gels11050335.
Postmastectomy radiation therapy (PMRT) is an adjuvant treatment for breast cancer. Some mastectomized women undergoing PMRT can have breast reconstruction with expander implant reconstruction. However, the expander implant contains a magnetic metal port for its inflation, and in patients with a high risk of recurrence, the PMRT is performed before the expander replacement. The difficulties in radiation treatment near high-Z metals are mainly due to dose alterations around them. Therefore, this study proposes using a realistic breast phantom and gel dosimetry to investigate the effects of the metallic parts of the expandable prosthesis on the 3D delivery of the treatment. A conformal radiation treatment was planned and delivered to the gel phantom with the metal port. MAGIC-f gel was used with magnetic resonance imaging for dose assessment. The treatment plan dose distribution was compared to the measured dose distribution by gamma analysis (3%/3 mm/15% threshold). A significant gamma fail region was found near the metal port, corresponding to a dose reduction of approximately 5%. This underdose is within the tolerance threshold for dose heterogeneity established by the International Commission on Radiation Units (ICRU), but should be considered when treating these patients.
乳房切除术后放射治疗(PMRT)是乳腺癌的一种辅助治疗方法。一些接受PMRT的乳房切除女性可以进行扩张器植入重建乳房。然而,扩张器植入物有一个用于充气的磁性金属端口,对于复发风险高的患者,在更换扩张器之前进行PMRT。在高原子序数金属附近进行放射治疗的困难主要是由于其周围的剂量改变。因此,本研究建议使用逼真的乳房模型和凝胶剂量测定法来研究可扩张假体的金属部件对治疗三维剂量传递的影响。计划对带有金属端口的凝胶模型进行适形放射治疗并实施。使用MAGIC-f凝胶并结合磁共振成像进行剂量评估。通过伽马分析(3%/3毫米/15%阈值)将治疗计划剂量分布与测量的剂量分布进行比较。在金属端口附近发现了一个显著的伽马失败区域,对应于约5%的剂量降低。这种剂量不足在国际辐射单位委员会(ICRU)确定的剂量异质性耐受阈值范围内,但在治疗这些患者时应予以考虑。