Pasler Marlies, Michel Kilian, Marrazzo Livia, Obenland Michael, Pallotta Stefania, Björnsgard Mari, Lutterbach Johannes
Lake Constance Radiation Oncology Center, Singen, Friedrichshafen, Germany.
Phys Med Biol. 2017 Sep 1;62(18):7440-7450. doi: 10.1088/1361-6560/aa7dc7.
The purpose of this study was to characterize a new single large-area ionization chamber, the integral quality monitor system (iRT, Germany), for online and real-time beam monitoring. Signal stability, monitor unit (MU) linearity and dose rate dependence were investigated for static and arc deliveries and compared to independent ionization chamber measurements. The dose verification capability of the transmission detector system was evaluated by comparing calculated and measured detector signals for 15 volumetric modulated arc therapy plans. The error detection sensitivity was tested by introducing MLC position and linac output errors. Deviations in dose distributions between the original and error-induced plans were compared in terms of detector signal deviation, dose-volume histogram (DVH) metrics and 2D γ-evaluation (2%/2 mm and 3%/3 mm). The detector signal is linearly dependent on linac output and shows negligible (<0.4%) dose rate dependence up to 460 MU min. Signal stability is within 1% for cumulative detector output; substantial variations were observed for the segment-by-segment signal. Calculated versus measured cumulative signal deviations ranged from -0.16%-2.25%. DVH, mean 2D γ-value and detector signal evaluations showed increasing deviations with regard to the respective reference with growing MLC and dose output errors; good correlation between DVH metrics and detector signal deviation was found (e.g. PTV D : R = 0.97). Positional MLC errors of 1 mm and errors in linac output of 2% were identified with the transmission detector system. The extensive tests performed in this investigation show that the new transmission detector provides a stable and sensitive cumulative signal output and is suitable for beam monitoring during patient treatment.
本研究的目的是对一种新型大面积电离室——整体质量监测系统(iRT,德国)进行特性描述,用于在线和实时束流监测。研究了静态和弧形照射时的信号稳定性、监测单位(MU)线性度和剂量率依赖性,并与独立电离室测量结果进行比较。通过比较15个容积调强弧形治疗计划的计算和测量探测器信号,评估了传输探测器系统的剂量验证能力。通过引入多叶准直器(MLC)位置和直线加速器输出误差来测试误差检测灵敏度。根据探测器信号偏差、剂量体积直方图(DVH)指标和二维γ评估(2%/2毫米和3%/3毫米),比较了原始计划和误差诱导计划之间的剂量分布偏差。探测器信号与直线加速器输出呈线性相关,在高达460 MU/min的剂量率范围内,剂量率依赖性可忽略不计(<0.4%)。累积探测器输出的信号稳定性在1%以内;逐段信号观察到显著变化。计算值与测量值的累积信号偏差范围为-0.16%至2.25%。DVH、平均二维γ值和探测器信号评估显示,随着MLC和剂量输出误差的增加,与各自参考值的偏差也在增加;发现DVH指标与探测器信号偏差之间具有良好的相关性(例如,计划靶体积(PTV)剂量:相关系数R = 0.97)。传输探测器系统识别出1毫米的MLC位置误差和2%的直线加速器输出误差。本研究中进行的广泛测试表明,新型传输探测器提供了稳定且灵敏的累积信号输出,适用于患者治疗期间的束流监测。