Med Eng Phys. 2017 Sep;47:159-166. doi: 10.1016/j.medengphy.2017.06.005. Epub 2017 Jul 3.
A new 2-D radiographic wear measurement system has been developed which enables the low wear of highly cross-linked polyethylene acetabular cups to be accurately and precisely measured from standard, pelvis radiographs. The software was validated using radiographic images of a measurement jig which could vary the cup orientation and simulate the effect of pelvic tilt/rotation. Wear was simulated using accurately measured plastic shims to vary the position of the femoral head relative to the cup. The effects of varying "wear" penetration, "wear" direction, cup orientation and X-ray focus position were assessed. Further direct comparison tests were also carried out using radiostereometric analysis. Inter/intra-observer repeatability of the new system was assessed using clinical radiographs. The mean (SD) "wear" penetration error was -0.002mm (0.028mm). The "wear" penetration precision was 0.055mm. Changing the position of the X-ray focus point made no difference to the measurement error. Inter/intra-observer repeatability and limits of agreement had similar low values. Comparison tests with RSA showed the accuracy was similar.
一种新的二维放射磨损测量系统已经开发出来,它能够从标准骨盆 X 光片中准确、精确地测量高度交联聚乙烯髋臼杯的低磨损。该软件使用测量夹具的放射图像进行了验证,该夹具可以改变杯的方向并模拟骨盆倾斜/旋转的影响。使用精确测量的塑料垫片模拟磨损,以改变股骨头相对于杯的位置。评估了“磨损”穿透、“磨损”方向、杯方向和 X 射线焦点位置的变化。还使用放射立体测量分析进行了进一步的直接比较测试。使用临床 X 光片评估了新系统的观察者内/间重复性。平均(标准差)“磨损”穿透误差为-0.002mm(0.028mm)。“磨损”穿透精度为 0.055mm。改变 X 射线焦点位置对测量误差没有影响。观察者内/间重复性和一致性界限也具有类似的低值。与 RSA 的比较测试表明,准确性相似。