Clin Orthop Relat Res. 2013 Jul;471(7):2238-44. doi: 10.1007/s11999-013-2789-x. Epub 2013 Jan 19.
One cross-linked polyethylene (XLPE) liner is manufactured using a lower dose of radiation, 5 Mrad, which may result in less cross-linking. The reported in vivo wear rate of this XLPE liner in patients undergoing THA has varied, and has included some patients in each reported cohort who had greater than 0.1 mm/year of wear, which is an historical threshold for osteolysis. Previous studies have measured wear on plain radiographs, an approach that has limited sensitivity.
QUESTIONS/PURPOSES: We therefore measured the amount and direction of wear at 6 years using Radiostereometric analysis (RSA) in patients who had THAs that included a cross-linked polyethylene liner manufactured using 5 Mrad radiation.
We prospectively reviewed wear in 30 patients who underwent primary THAs with the same design of cross-linked acetabular liner and a 28-mm articulation. Tantalum markers were inserted during surgery and all patients had RSA radiographic examinations at 1 week, 6 months, 1, 2, and 6 years postoperatively.
The mean proximal, two-dimensional (2-D) and three-dimensional (3-D) wear rates calculated between 1 year and 6 years were 0.014, 0.014, and 0.018 mm/per year, respectively. The direction of the head penetration recorded between 1 week and 6 years was in a proximal direction for all patients, proximolateral for 16 of 24 patients, and proximomedial for eight of 24 patients.
The proximal, 2-D and 3-D wear of a XLPE liner produced using 5 Mrad of radiation was low but measurable by RSA after 6 years. No patients had proximal 2-D or 3-D wear rates exceeding 0.1 mm/year. Further followup is needed to evaluate the effect of XLPE wear particles on the development of long-term osteolysis.
一种交联聚乙烯(XLPE)衬垫采用较低剂量的辐射(5 Mrad)制造,这可能导致交联程度降低。接受全髋关节置换术(THA)的患者中,这种 XLPE 衬垫的报告体内磨损率各不相同,每个报告队列中都有一些患者的磨损率超过 0.1 毫米/年,这是骨溶解的历史阈值。以前的研究使用普通射线照相术测量磨损,这种方法的灵敏度有限。
问题/目的:因此,我们使用放射立体测量分析(RSA)测量了接受使用 5 Mrad 辐射制造的交联聚乙烯衬垫的 THA 的 6 年后的磨损量和方向。
我们前瞻性地回顾了 30 名接受初次 THA 的患者的磨损情况,这些患者采用相同设计的交联髋臼衬垫和 28 毫米关节。手术期间插入钽标记物,所有患者在术后 1 周、6 个月、1 年、2 年和 6 年接受 RSA 放射照相检查。
在 1 年至 6 年之间计算的近端、二维(2-D)和三维(3-D)磨损率平均值分别为 0.014、0.014 和 0.018 毫米/年。在 1 周到 6 年之间记录的头穿透方向对于所有患者均为近端,对于 24 名患者中的 16 名患者为近侧前外侧,对于 24 名患者中的 8 名患者为近侧内侧。
使用 5 Mrad 辐射制造的 XLPE 衬垫的近端、2-D 和 3-D 磨损率较低,但在 6 年后通过 RSA 可测量。没有患者的近端 2-D 或 3-D 磨损率超过 0.1 毫米/年。需要进一步随访以评估 XLPE 磨损颗粒对长期骨溶解发展的影响。