Fahey Frederic H, Goodkind Alison, MacDougall Robert D, Oberg Leah, Ziniel Sonja I, Cappock Richard, Callahan Michael J, Kwatra Neha, Treves S Ted, Voss Stephan D
Division of Nuclear Medicine and Molecular Imaging, Department of Radiology, Boston Children's Hospital, Boston, Massachusetts
Department of Radiology, Harvard Medical School, Boston, Massachusetts.
J Nucl Med. 2017 Sep;58(9):1360-1366. doi: 10.2967/jnumed.116.182899. Epub 2017 Jul 7.
No consistent guidelines exist for the acquisition of a CT scan as part of pediatric PET/CT. Given that children may be more vulnerable to the effects of ionizing radiation, it is necessary to develop methods that provide diagnostic-quality imaging when needed, in the shortest time and with the lowest patient radiation exposure. This article describes the basics of CT dosimetry and PET/CT acquisition in children. We describe the variability in pediatric PET/CT techniques, based on a survey of 19 PET/CT pediatric institutions in North America. The results of the survey demonstrated that, although most institutions used automatic tube current modulation, there remained a large variation of practice, on the order of a factor of 2-3, across sites, pointing to the need for guidelines. We introduce the approach developed at our institution for using a multiseries PET/CT acquisition technique that combines diagnostic-quality CT in the essential portion of the field of view and a low-dose technique to image the remainder of the body. This approach leads to a reduction in radiation dose to the patient while combining the PET and the diagnostic CT into a single acquisition. The standardization of pediatric PET/CT provides an opportunity for a reduction in the radiation dose to these patients while maintaining an appropriate level of diagnostic image quality.
作为儿科PET/CT的一部分,目前尚无关于获取CT扫描的统一指南。鉴于儿童可能更容易受到电离辐射的影响,有必要开发在需要时能在最短时间内以最低的患者辐射暴露提供诊断质量成像的方法。本文介绍了儿童CT剂量测定和PET/CT采集的基础知识。我们基于对北美19家儿科PET/CT机构的调查,描述了儿科PET/CT技术的变异性。调查结果表明,尽管大多数机构使用自动管电流调制,但各机构之间在实践上仍存在很大差异,相差2至3倍,这表明需要制定指南。我们介绍了我们机构开发的一种方法,即使用多系列PET/CT采集技术,该技术在视野的关键部分结合诊断质量的CT,并使用低剂量技术对身体的其余部分进行成像。这种方法在将PET和诊断CT合并为一次采集的同时,降低了患者的辐射剂量。儿科PET/CT的标准化为降低这些患者的辐射剂量提供了机会,同时保持适当水平的诊断图像质量。