Handheld fetal Doppler device for assessing heart rate in neonatal resuscitation.

作者信息

Shimabukuro Rinshu, Takase Kozo, Ohde Sachiko, Kusakawa Isao

机构信息

Department of Pediatrics, St Luke's International Hospital, Tokyo, Japan.

Department of Research Development, Division of Public Health, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University, Tokyo, Japan.

出版信息

Pediatr Int. 2017 Oct;59(10):1069-1073. doi: 10.1111/ped.13374. Epub 2017 Oct 5.

Abstract

BACKGROUND

The use of neonatal three-lead electrocardiograms (Neo-ECG) is suggested for the assessment of heart rate (HR) in neonatal resuscitation, but primary childbirth facilities (i.e. maternity clinics or maternity homes) are not sufficiently equipped for this. The aim of this study was to therefore to confirm the potential of a handheld fetal Doppler device (FDD), widely used to assess HR in labor room resuscitation, and to compare its equivalency to Neo-ECG.

METHODS

In low-risk full-term infants, HR was measured five times per minute in minutes 1-5 after delivery using the FDD and Neo-ECG, and their equivalence compared. The equivalence margin was set at a HR difference of <±5 beats/min and <±5%. Moreover, the times at which HR was displayed were measured for both methods.

RESULTS

FDD and Neo-ECG HR correlated strongly (r = 0.977) in 155 measurements in 33 cases. The mean HR difference was -0.123 beats/min (95%CI: -0.594 to +0.345), and the logarithmic transformation of the HR ratio was -0.0003 (95%CI: -0.0020 to +0.0015). The mean HR difference and HR ratio were within the equivalence margin. For HR assessment using the FDD, 155 of 165 measurements (93.9%) could be obtained in <10 s.

CONCLUSIONS

In HR assessment, FDD is a valid substitute for Neo-ECG in primary childbirth facilities that are not sufficiently equipped for Neo-ECG.