J Matern Fetal Neonatal Med. 2021 Jul;34(13):2053-2060. doi: 10.1080/14767058.2019.1656193. Epub 2019 Aug 27.
The assessment of newborns' heart rate (HR) in the delivery room is one of the important steps to ascertain the need for initiation and continuation of resuscitation. At present, ECG is the "gold standard" to monitor neonatal HR in the delivery room. However, various limitations with the use of ECG exist. Furthermore, in developing countries, ECG may not be universally available in delivery rooms.
To compare the accuracy and HR acquisition time of portable Doppler ultrasound (PDU) versus electrocardiogram (ECG) in newborns.
This multicenter, prospective, observational study across five centers in India between January and September 2017 included neonates more than 34 weeks of gestation ( = 131) delivered by cesarean section. The accuracy of HR recorded by PDU (HR) versus that by ECG (HR) was the primary outcome. Secondary outcomes included time to acquisition of an audible and/or visible signal and device application.
Mean (±SD) gestational age and birthweight were 37.7 (±1.2) weeks and 2954 (±457) g, respectively. The mean (±SD) visible HR was 158 (±21) bpm versus HR of 161.3 (±20) bpm ( = .07) which were comparable. The median (1st, third quartile) time to acquisition of audible HR (76 [51, 91] s), was significantly shorter than that of HR (96.5 [74.2, 118] s; < .001).
Portable Doppler has similar accuracy to ECG and is faster in acquiring the signal.
评估新生儿在产房的心率(HR)是确定是否需要开始和继续复苏的重要步骤之一。目前,心电图(ECG)是产房监测新生儿 HR 的“金标准”。然而,ECG 的使用存在各种局限性。此外,在发展中国家,心电图可能并非普遍适用于产房。
比较便携式多普勒超声(PDU)与心电图(ECG)在新生儿中的准确性和 HR 获取时间。
这是一项在印度五个中心进行的多中心、前瞻性、观察性研究,于 2017 年 1 月至 9 月期间纳入了 34 周以上剖宫产分娩的新生儿(n=131)。PDU 记录的 HR(HR)与 ECG 记录的 HR(HR)的准确性是主要结局。次要结局包括获取可听和/或可见信号的时间以及设备应用。
平均(±SD)胎龄和出生体重分别为 37.7(±1.2)周和 2954(±457)g。平均(±SD)可见 HR 为 158(±21)bpm,而 HR 为 161.3(±20)bpm( = .07),两者相当。获取可听 HR 的中位数(1 四分位数,3 四分位数)时间(76[51,91]s)明显短于 HR 的时间(96.5[74.2,118]s; < .001)。
便携式多普勒与 ECG 具有相似的准确性,并且在获取信号方面更快。