Chiu W Brian, de Bie Johan, Mortara David W
Mortara Instrument, Inc., 7865 North 86th St., Milwaukee, WI 53204, USA.
Mortara Instrument, Inc., Via Cimarosa 103, 40033 Casalecchio di Reno, BO, Italy.
J Electrocardiol. 2017 Nov-Dec;50(6):758-761. doi: 10.1016/j.jelectrocard.2017.07.011. Epub 2017 Jul 12.
To evaluate performance of J-to-T-peak (JTP) measurements of 12-lead ECGs, in a five-arm study using drugs with various levels of electrolyte channel block.
The novel evaluation method distinguishes between different aspects of measurement. "Random noise" is the variability among repeated measurements made without changing the conditions. "Context noise" is the variability of changes in context of the measurement, e.g. T-wave morphology, autonomic nervous system state.
The average random noise of our RR-corrected JTPc measurements in standard deviations was 3.0 ms and not dependent on the drug. The average context noise was 4.0 ms for ranolazine, verapamil, and placebo, and 8.8 ms for dofetilide and quinidine. Measurement consistency is corroborated by linear fit confidence intervals of baseline- and placebo-corrected JTPc versus drug concentration.
Systematic differences were found in JTPc drug response between the Mortara method and published data. Residual signal component in the context noise may influence future study design.