Singh S, Slavicek T, Hodak R, Versaci R, Pridal P, Kumar D
Institute of Physics of the ASCR, ELI-Beamlines, 18221 Prague, Czech Republic.
Institute of Experimental and Applied Physics, Czech Technical University in Prague, Prague, Czech Republic.
Rev Sci Instrum. 2017 Jul;88(7):075105. doi: 10.1063/1.4993921.
This paper presents the calibration of two different kinds of image plates (IPs) for detecting electrons with kinetic energy in the range of 150 keV-1.75 MeV. The calibration was performed using a Sr β source. The paper also provides the measured fading response for the IPs in the time range from 12 min to 18 h. Calibration results are compared to Monte Carlo simulations of energy deposited by the electrons in the sensitive layer of the IPs. It was found that within this energy range a linear relation between simulated energy deposited by the electron in the phosphor layer and the measured photo stimulated luminescence in the IP is adequate to model the response of the IP.