Wesselowski S, Saunders A B, Gordon S G
College of Veterinary Medicine, Texas A&M University, College Station, TX.
J Vet Intern Med. 2017 Sep;31(5):1388-1391. doi: 10.1111/jvim.14797. Epub 2017 Aug 7.
Deployment of the Amplatz Canine Duct Occluder (ACDO) is the preferred method for minimally invasive occlusion of patent ductus arteriosus (PDA) in dogs, with appropriate device sizing crucial to successful closure. Dogs of any body weight can be affected by PDA.
To describe the range of ACDO sizes deployed in dogs of various body weights for improved procedural planning and inventory selection and to investigate for correlation between minimal ductal diameter (MDD) and body weight.
A total of 152 dogs undergoing ACDO deployment between 2008 and 2016.
Body weight, age, breed, sex, and MDD obtained by angiography (MDD-A), MDD obtained by transesophageal echocardiography (MDD-TEE), and ACDO size deployed were retrospectively evaluated.
Correlation between body weight and ACDO size, MDD-A and MDD-TEE was poor, with R-squared values of 0.4, 0.36, and 0.3, respectively. Femoral artery diameter in the smallest population of dogs placed inherent limitations on the use of larger device sizes, with no limitations on the wide range of device sizes required as patient size increased. The most commonly used ACDO devices were size 3 through 6, representing 57% of the devices deployed within the entire study population.
Patent ductus arteriosus anatomy varies on an individual basis, with poor correlation between MDD and body weight. Weight-based assumptions about expected ACDO device size for a given patient are not recommended.
使用Amplatz犬用动脉导管封堵器(ACDO)是犬动脉导管未闭(PDA)微创封堵的首选方法,合适的器械尺寸对于成功封堵至关重要。任何体重的犬都可能患动脉导管未闭。
描述在不同体重犬中使用的ACDO尺寸范围,以改进手术规划和库存选择,并研究最小导管直径(MDD)与体重之间的相关性。
2008年至2016年间共有152只接受ACDO植入的犬。
回顾性评估体重、年龄、品种、性别,以及通过血管造影获得的MDD(MDD-A)、经食管超声心动图获得的MDD(MDD-TEE)和所使用的ACDO尺寸。
体重与ACDO尺寸、MDD-A与MDD-TEE之间的相关性较差,决定系数分别为0.4、0.36和0.3。最小犬群体的股动脉直径对使用更大尺寸器械存在固有限制,而随着患者体型增加,所需器械尺寸范围广泛则没有限制。最常用的ACDO器械尺寸为3至6号,占整个研究群体中植入器械的57%。
动脉导管未闭的解剖结构因个体而异,MDD与体重之间相关性较差。不建议基于体重假设给定患者预期的ACDO器械尺寸。