Shurrab Mohammed, Pagacz Teresa, Shauer Ayelet, Lashevsky Ilan, Newman David, Crystal Eugene
Arrhythmia Services, Schulich Heart Centre, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, ON, Canada.
Division of Cardiology, Women's College Hospital, University of Toronto, Toronto, ON, Canada.
Clin Med Insights Cardiol. 2017 Jun 23;11:1179546817714478. doi: 10.1177/1179546817714478. eCollection 2017.
Ventricular safety pacing (VSP) is used to avoid cross talk by delivering ventricular stimulus shortly after an atrial-paced event if ventricular-sensed event occurs. Although VSP is a protective feature that exists for decades in different pacing devices, there are some reports of unfavorable outcomes of this algorithm. More so, health care providers sometimes face difficulties in interpreting and dealing with VSP strips. This case report discusses an important pacemaker algorithm and encourages further attention to possible pitfalls and hence avoids unnecessary interventions.
心室安全起搏(VSP)用于在心房起搏事件后若出现心室感知事件时,通过快速发放心室刺激来避免串扰。尽管VSP作为一项保护功能已在不同起搏设备中存在数十年,但仍有一些关于该算法不良后果的报告。更有甚者,医疗保健人员有时在解读和处理VSP心电图记录条时会遇到困难。本病例报告讨论了一种重要的起搏器算法,并鼓励进一步关注可能存在的隐患,从而避免不必要的干预。