In Vitro Evaluation of the Occlusive Properties of the ArtVentive Endoluminal Occlusion System Occlusion Device for Transrenal Ureteral Occlusion.

作者信息

Kuetting Daniel, Meyer Carsten, Schild Hans Heinz, Pieper Claus Christian

机构信息

Department of Radiology, University of Bonn , Bonn, Germany .

出版信息

J Endourol. 2017 Oct;31(10):1084-1089. doi: 10.1089/end.2017.0493. Epub 2017 Sep 14.

Abstract

PURPOSE

Current techniques for percutaneous ureteral occlusion are either technically difficult or not satisfactory because of frequent ureteral recanalization. The purpose of this in vitro study was to evaluate the occlusive properties of an "off the shelf" solution (Endoluminal Occlusion System [EOS™]; ArtVentive Medical Group, Inc., Carlsbad, CA) for transrenal ureteral occlusion.

MATERIALS AND METHODS

Both 8 and 11 mm expanded polytetrafluoroethylene-covered ArtVentive EOS devices were used in 10 porcine models. Experiments were performed in explanted porcine ureters to simulate physiologic conditions. EOS devices were deployed in a midureteral position using a transrenal approach. Contrast agent (Iopamidol 300) diluted in saline solution was infused into the renal pelvis under continuous fluoroscopic guidance. Intrapelvic pressure measurements were performed until leakage, plug dislocation, or until pelvic blow out occurred.

RESULTS

All EOS devices were deployed effectively and achieved prompt total ureteral occlusion. Ureteral leakage occurred with intraureteral pressures between 60 to 109 cm HO (8 mm EOS) and between 65 and 125 cm HO (11 mm EOS). Before leakage, tubular reflux was seen in all cases, pelvic blowout occurred in half of the cases.

CONCLUSIONS

The ArtVentive EOS occlusive device is an effective tool for "off the shelf" ureteral occlusion. Both the 8 mm and the 11 mm devices fully occluded ureters at pressure levels that are to be expected in vivo.

摘要

目的

由于输尿管再通频繁,目前经皮输尿管闭塞技术要么操作困难,要么效果不理想。本体外研究的目的是评估一种“现成可用”的溶液(腔内闭塞系统[EOS™];ArtVentive Medical Group, Inc., 加利福尼亚州卡尔斯巴德)用于经肾输尿管闭塞的闭塞特性。

材料与方法

在10只猪模型中使用了8毫米和11毫米的膨体聚四氟乙烯覆盖的ArtVentive EOS装置。在离体猪输尿管中进行实验以模拟生理条件。使用经肾途径将EOS装置部署在输尿管中段位置。在连续透视引导下,将稀释于盐溶液中的造影剂(碘帕醇300)注入肾盂。进行肾盂内压力测量,直至出现渗漏、封堵器移位或肾盂破裂。

结果

所有EOS装置均有效部署并迅速实现完全输尿管闭塞。输尿管渗漏发生时,输尿管内压力在60至109厘米水柱之间(8毫米EOS)以及65至125厘米水柱之间(11毫米EOS)。在渗漏前,所有病例均可见肾小管反流,半数病例发生肾盂破裂。

结论

ArtVentive EOS闭塞装置是一种用于“现成可用”输尿管闭塞的有效工具。8毫米和11毫米装置在体内预期的压力水平下均能完全闭塞输尿管。