Suppr超能文献

一种用于经心尖进入主动脉瓣、二尖瓣以及主动脉各节段的新型实验装置。

A New Experimental Device for Transapical Access of the Aortic and Mitral Valves as well as the Aorta in its Various Segments.

作者信息

Paim Leonardo, Fonseca José Honório Palma da, Arruda Francismar Vidal de, Gutierrez Paulo Sampaio, Moreira Luiz Felipe Pinho, Jatene Fabio Biscegli

机构信息

Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo (InCor-HCFMUSP), São Paulo, SP, Brazil.

出版信息

Braz J Cardiovasc Surg. 2017 May-Jun;32(3):156-161. doi: 10.21470/1678-9741-2017-0013.

Abstract

OBJECTIVE

: To present the results of a new experimental device developed to facilitate the transapical access in endovascular treatment of structural heart diseases. It aims to reduce the risk of bleeding and complications in this type of access and demonstrate the device as a safe, fast and effective alternative.

METHODS

: CorPoint is composed of three parts: introducer, base with coiled spring, and closing capsule. By rotating movements, the spring is introduced into the myocardium and progressively approaches the base to the surface of the heart. Guidewires and catheters are inserted through the hollow central part and, at the end of the procedure, the capsule is screwed over the base, therefore stopping any bleeding.

RESULTS

: The device was implanted in 15 pigs, weighing 60 kg each, through an anterolateral thoracotomy, while catheters were introduced and guided by fluoroscopy. All animals had minimal bleeding; introducers with diameter up to 22 Fr were used and various catheters and guidewires were easily handled. After finishing the procedure, the closing capsule was attached and no bleeding was observed at the site.

CONCLUSION

: This new device has proved effective, fast and secure for the transapical access. This shows great potential for use, especially by ensuring an easier and direct access to the mitral and aortic valves; the shortest distance to be traveled by catheters; access to the ascending and descending aorta; decreased bleeding complications; decreased surgical time; and the possibility of allowing the technique to evolve and become totally percutaneous.

摘要

目的

展示一种新开发的实验装置的结果,该装置旨在促进结构性心脏病血管内治疗中的经心尖入路。其目的是降低这种入路方式的出血和并发症风险,并证明该装置是一种安全、快速且有效的替代方法。

方法

CorPoint由三部分组成:导入器、带螺旋弹簧的基座和封闭胶囊。通过旋转动作,将弹簧引入心肌,并使基座逐渐靠近心脏表面。导丝和导管通过中空的中心部分插入,在操作结束时,将胶囊拧在基座上,从而阻止任何出血。

结果

通过前外侧开胸手术,将该装置植入15头体重均为60千克的猪体内,同时在荧光透视引导下插入导管。所有动物出血极少;使用了直径达22F的导入器,各种导管和导丝都易于操作。操作完成后,安装封闭胶囊,在该部位未观察到出血。

结论

这种新装置已被证明在经心尖入路方面有效、快速且安全。这显示出巨大的应用潜力,特别是通过确保更简便、直接地进入二尖瓣和主动脉瓣;导管行进的最短距离;进入升主动脉和降主动脉;减少出血并发症;缩短手术时间;以及使该技术有可能发展并完全实现经皮操作。