在体外,使用syngo iFlow基于造影剂评估血流分流器大小和位置对动脉瘤内血流动力学的影响。

In vitro, contrast agent-based evaluation of the influence of flow diverter size and position on intra-aneurysmal flow dynamics using syngo iFlow.

作者信息

Cattaneo Giorgio Franco Maria, Ding Andreas, Jost Tobias, Ley Désirée, Mühl-Bennighaus Ruben, Yilmaz Umut, Körner Heiko, Reith Wolfgang, Simgen Andreas

机构信息

Acandis GmbH, Theodor-Fahrner-Str. 6, 75177, Pforzheim, Germany.

Departments of Neuroradiology, Saarland University Hospital, Homburg, Saar, Germany.

出版信息

Neuroradiology. 2017 Dec;59(12):1275-1283. doi: 10.1007/s00234-017-1903-2. Epub 2017 Aug 25.

Abstract

PURPOSE

Treatment of intracranial aneurysm with flow-diverting devices has become widespread in recent years. Despite that, intra-aneurysmal flow changes are yet not fully understood and can lead to different complications. Our aim was an in vitro contrast-based evaluation of the influence of flow diverter size and position on intra-aneurysmal flow dynamics.

METHODS

Flow-diverting devices with different sizes (diameters 4.0, 4.5, and 6.0 mm) were deployed in seven silicone aneurysm models at different positions relative to the aneurysm neck (proximal, central, distal). Using syngo iFlow, we defined quantitative evaluation criteria based on contrast medium intensity and performed a flow evaluation.

RESULTS

Intra-aneurysmal flows were heavily dependent on both size and position of flow-diverting devices at the aneurysm neck. We observed a higher peak intensity delay and intra-aneurysmal washout delay with the centrally placed 4.0- and 4.5-mm device, respectively, compared to the proximal and distal positions. Especially distally placed 4.0-mm devices led to an earlier filling of the aneurysm and increased intra-aneurysmal contrast agent intensity compared to the parent vessel, due to a potential endoleak.

CONCLUSIONS

Not only size but also position of flow-diverting devices have a considerable impact on the intra-aneurysmal flow dynamics. The suggested evaluation criteria allowed a quantitative comparison of flow-diverting effect using syngo iFlow and could represent an efficient tool for predicting flow diversion pre-procedurally.

摘要

目的

近年来,使用血流导向装置治疗颅内动脉瘤已变得广泛。尽管如此,动脉瘤内的血流变化尚未完全了解,并且可能导致不同的并发症。我们的目的是基于造影剂对血流导向装置的尺寸和位置对动脉瘤内血流动力学的影响进行体外评估。

方法

将不同尺寸(直径4.0、4.5和6.0毫米)的血流导向装置部署在七个硅胶动脉瘤模型中,相对于动脉瘤颈部处于不同位置(近端、中央、远端)。使用syngo iFlow,我们基于造影剂强度定义了定量评估标准并进行了血流评估。

结果

动脉瘤内的血流严重依赖于动脉瘤颈部血流导向装置的尺寸和位置。我们观察到,与近端和远端位置相比,置于中央的4.0毫米和4.5毫米装置分别具有更高的峰值强度延迟和动脉瘤内洗脱延迟。特别是,由于可能存在内漏,置于远端的4.0毫米装置导致动脉瘤更早充盈,并且与母血管相比,动脉瘤内造影剂强度增加。

结论

血流导向装置的尺寸和位置不仅对动脉瘤内血流动力学有相当大的影响。所建议的评估标准允许使用syngo iFlow对血流导向效果进行定量比较,并且可以代表一种在术前预测血流导向的有效工具。