Relationship Between Pump Speed Design and Hemolysis in an Axial Flow Blood Pump.

作者信息

Anai Hirofumi, Wakisaka Yoshinari, Nakatani Takeshi, Taenaka Yoshiyuki, Takano Hisateru, Hadama Tetsuo

机构信息

Texas Heart Institute, Houston, Texas, U.S.A.National Cardiovascular Center, OsakaOita Medical University, Oita, Japan.

出版信息

Artif Organs. 1996 May;20(5):564-567. doi: 10.1111/j.1525-1594.1996.tb04481.x.

Abstract

In an attempt to reduce the hemolysis caused by axial flow blood pumps, we investigated whether the specific speed should be kept within the standard engineering range or whether pump speed should be minimized, thus making the specific speed beyond the standard range. Four pumps with 11.8 mm diameter impellers were designed to accommodate a flow of 5 L/min and a head of 100 mm Hg. The pumps were tested at 4 speeds: A, 14,000; B, 18,000; C, 22,000; and D, 26,000 rpm. Pump performance data were obtained, and the maximum point of total pump efficiency was found for each pump. The maximum efficiencies were A, 50%; B, 58%; C, 52%; and D, 53%. The specific speed of each pump recorded at the maximum efficiency point was calculated as A, 899; B, 954; C, 1,218; and D, 1,951 rpm. Hemolytic tests were performed with fresh goat blood in a closed, mock-loop circuit. Hemolytic indexes were A, 0.036; B, 0.22; C, 0.35; and D, 0.66. We have concluded that decreased hemolysis is correlated with a lower pump speed and that the specific speed for the lowest pump speed is less than the standard range. Having a specific speed outside the standard range was not correlated with reduced total pump efficiency.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验