Emmace Consulting AB, SE-22381, Lund, Sweden.
Jolyon Mitchell Inhaler Consulting Services Inc., 1154 St. Anthony Road, London, Ontario, N6H 2R1, Canada.
AAPS PharmSciTech. 2018 Feb;19(2):565-572. doi: 10.1208/s12249-017-0860-8. Epub 2017 Sep 5.
Determination of fine droplet dose with preparations for nebulization, currently deemed to be the metric most indicative of lung deposition and thus in vivo responses, involves combining two procedures following practice as described in the United States Pharmacopeia and the European Pharmacopeia. Delivered dose (DD) is established by simulating tidal breathing at the nebulizer, collecting the medication on a filter downstream of the nebulizer mouthpiece/facemask. Fine droplet fraction (FDF) is determined separately using a cooled cascade impactor operated at 15 L/min. FDD is subsequently calculated as the product of DD and FDF. Development of the Nephele mixing inlet has allowed cascade impactor-based assessments to be made at a constant flow rate while simultaneously subjecting the nebulizer to the continuously varying flow profile associated with breath simulation. The study purpose was to investigate the feasibility of this approach, termed mixing inlet lung simulation (MILS), for direct determination of FDD. An optimal upper size limit for FDF is not given for nebulizers, but 5 μm was chosen since this limit is the European norm when testing other inhalation products. Vibrating membrane nebulizers (eFlow® Rapid) were used to deliver aqueous salbutamol sulfate, simulating an adult tidal-breathing pattern (inspiratory to expiratory ratio = 1:1, tidal volume = 500 mL, 15 breaths per minute, peak inspiratory flow rate = 24 L/min). The two procedures were inequivalent, as FDD by the MILS approach was 72% of that obtained using the compendial "combination" method. Since the MILS methodology more closely mimics clinical use, we infer that the compendial approach likely overestimates the dose reaching the human lung.
确定雾化制剂的微细液滴剂量,目前被认为是最能反映肺部沉积和体内反应的指标,需要结合美国药典和欧洲药典中描述的两种方法来进行。通过在雾化器上模拟潮式呼吸,在雾化器喷口/面罩下游的过滤器上收集药物,从而确定输送剂量(DD)。使用在 15 L/min 下运行的冷却级联冲击器分别确定微细液滴分数(FDF)。随后将 DD 和 FDF 的乘积作为 FDD 进行计算。Nephele 混合入口的开发允许在恒流率下进行基于级联冲击器的评估,同时使雾化器经受与呼吸模拟相关的不断变化的流量分布。该研究的目的是调查这种称为混合入口肺模拟(MILS)的方法对于直接确定 FDD 的可行性。对于雾化器,没有给出微细液滴分数的最佳上限,但选择 5μm,因为当测试其他吸入产品时,这是欧洲标准。使用振动膜雾化器(eFlow® Rapid)输送硫酸沙丁胺醇水溶液,模拟成人潮式呼吸模式(吸呼比=1:1,潮气量=500mL,每分钟 15 次呼吸,吸气峰流速=24L/min)。这两种方法是不等效的,因为 MILS 方法的 FDD 仅为采用药典“组合”方法获得的 FDD 的 72%。由于 MILS 方法更接近临床使用,我们推断药典方法可能高估了到达人体肺部的剂量。