Leblans Paul, Vandenbroucke Dirk, Willems Peter
Agfa HealthCare NV, Septestraat 27, B-2640 Mortsel, Belgium.
Industrial Scientific and Computing BVBA, Hellestraat 55, B-9190 Stekene, Belgium.
Materials (Basel). 2011 Jun 9;4(6):1034-1086. doi: 10.3390/ma4061034.
Computed radiography (CR) uses storage phosphor imaging plates for digital imaging. Absorbed X-ray energy is stored in crystal defects. In read-out the energy is set free as blue photons upon optical stimulation. In the 35 years of CR history, several storage phosphor families were investigated and developed. An explanation is given as to why some materials made it to the commercial stage, while others did not. The photo stimulated luminescence mechanism of the current commercial storage phosphors, BaFBr:Eu and CsBr:Eu is discussed. The relation between storage phosphor plate physical characteristics and image quality is explained. It is demonstrated that the morphology of the phosphor crystals in the CR imaging plate has a very significant impact on its performance.
计算机X线摄影(CR)使用存储磷光成像板进行数字成像。吸收的X射线能量存储在晶体缺陷中。在读取过程中,能量在光激发下以蓝色光子的形式释放出来。在CR的35年历史中,人们研究和开发了几种存储磷光体家族。文中解释了为什么有些材料进入了商业阶段,而有些则没有。讨论了当前商业存储磷光体BaFBr:Eu和CsBr:Eu的光激发发光机制。解释了存储磷光板物理特性与图像质量之间的关系。结果表明,CR成像板中磷光体晶体的形态对其性能有非常显著的影响。